80T-150UA Fluke Electronics, 80T-150UA Datasheet

no-image

80T-150UA

Manufacturer Part Number
80T-150UA
Description
UNIVERSAL TEMPERATURE PROBE
Manufacturer
Fluke Electronics
Type
Temperature Prober
Datasheet

Specifications of 80T-150UA

Usage
Gas, Liquid, Solid
Tip - Type
Contact, Spring
Plug Type
Banana Plug, Sheathed (2)
Probe Temperature Range
-58 ~ 302°F (-50 ~ 150°C)
Cable Length
46.000" (1168.40mm)
Plug Color
Black and Red
Probe Material
Valox, Glass Filled
For Use With
614-1036 - THERMTR CONTCT DUAL INP W/RECRD614-1035 - THERMTR CONTCT SNGL INP W/RECRD614-1034 - THERMTR CONTACT DUAL INPUT
Lead Free Status / RoHS Status
Lead free / Request inventory verification
Probe Length
-
Cable Insulation
-
Introduction
The Model 80T-150UA Universal Temperature Probe is a self-contained
temperature-to-voltage converter. The probe is designed to provide a
direct temperature reading when it is connected to any high impedance
DMM that is capable of 1 mV resolution and at least a 300-count full
scale readout capability. Output is 1 mV per degree (Celsius or
Fahrenheit). Two switch-selected temperature output scalings are
provided: -50 to +150 C or -58 to +302 F. The probe will stand off 60 V
dc or 30 V ac rms (42.4 V peak).
The unit is housed in two separate but attached assemblies: a
temperature probe and a temperature-to-voltage converter. The probe
contains the temperature-sensing element and is electrically connected to
the temperature-to-voltage converter through a 46-inch cable. A three-
position switch on the converter acts as a power switch and is used for
selecting Celsius or Fahrenheit scaling for the output. Two banana plugs
with standard 0.75-inch spacing are provided for connecting the 80T-
150UA to the DMM.
Operating power for the 80T-150UA is derived from a standard 9 V
battery. Typically, an alkaline battery provides more than 600 hours of
continuous operation before replacement is necessary. An OFF switch is
provided on the temperature-to-voltage converter to allow battery
conservation when the unit is not in use. In addition, the OFF position of
the power switch allows the battery condition to be determined via the
external DMM.
Temperature is measured by exposing the probe tip directly to the
material to be measured (non-corrosive liquid, gas, or solid). A direct
temperature reading is displayed on the DMM.
Symbols
PN 3208738
February 2008
© 2008 Fluke Corporation. All rights reserved.
All product names are trademarks of their respective companies.
XWarning
W Attention:
Universal Temperature Probe
:
Risk of electric shock. Refer to the manual for
information.
Refer to the manual for information about this
feature.
80T-150UA
Instruction Sheet
®

80T-150UA Summary of contents

Page 1

... Celsius or Fahrenheit scaling for the output. Two banana plugs with standard 0.75-inch spacing are provided for connecting the 80T- 150UA to the DMM. Operating power for the 80T-150UA is derived from a standard 9 V battery. Typically, an alkaline battery provides more than 600 hours of continuous operation before replacement is necessary. An OFF switch is provided on the temperature-to-voltage converter to allow battery conservation when the unit is not in use ...

Page 2

... Specifications The 80T-150UA will achieve rated accuracy when it is used with any 0.25 % DMM that has an input impedance Electrical Measurement Range: (See Probe Limitations) Accuracy: (Applies for one year after purchase or calibration) AMBIENT C + +15 C and +35 to +50 C AMBIENT F +59 to +95 F ...

Page 3

... The operator should read these paragraphs before attempting to operate the probe. Probe Limitations The 80T-150UA probe is constructed of a highly durable plastic and is suitable for measuring the temperature of liquids, gases, and solid surfaces up to 150 C. When measuring temperature, observe the following precautions to prevent damage to the probe: 1 ...

Page 4

... The 2 V range 1/2-digit DMM is adequate. Ignore readings of less than 1 when a more sensitive DMM is used. 3. Set the 80T-150UA power switch and energize the DMM. 4. Firmly touch the probe tip to the surface to be measured, or expose liquid or gas. The DMM will display the temperature in degrees ...

Page 5

... Return the probe tip to the 0 C bath wait 10 minutes, then check the output. If readjustment is necessary, repeat steps 4 through 8 until the readings can be obtained without adjustment. 10. Set the 80T-150UA switch to the OFF position, and remove the 80T- 150UA from the DVM. 11. Reassemble the 80T-150UA. ...

Page 6

... MP101 MP102 Figure 2. 80T-150UA Temperature Probe Instrument Type Mercury Thermometer Dewar Flask and Cap Metal or Glass Container Digital Voltmeter To contact Fluke or for service, call one of the following telephone numbers: USA: 1-888-44-FLUKE (1-888-443-5853), Canada: 1-800-36-FLUKE (1-800-363- 5853), Europe: +31 402-675-200, Japan: +81-3-3434-0181, Singapore: +65-738- 5655, Anywhere in the world: +1-425-446-5500 Or, visit Fluke's Web site at www ...