Specifications of FLUKE-52-2 60HZ

Temperature Range
-418 ~ 2501°F (-250 ~ 1372°C)
Temperature Display
C°/F°/K°
Display Type
LCD
Input Type
Thermocouples (2)
Features
Backlight, Hold, MIN/MAX/Avg, Sleep Mode
Probe Type
J, K, T, E
Battery Type
AA (3)
Includes
Holster, Software, Thermocouple - K-Type Bead Probe (2)
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Other names
52-2
52-2 60HZ
614-1034
674689
51 & 52 Series II
Users Manual
Holding the Displayed Readings
1.
2.
3.
Viewing the MIN, MAX, and AVG Readings
1.
2.
12
Press
Press
Press
The display shows
Model 52: Press
T1-T2 readings in the primary or secondary display.
Press
minimum (MIN), or the average (AVG) readings.
The elapsed time since entering MIN MAX mode, or
the time at which the minimum or maximum occurred,
appears on the display.
,
to step through the maximum (MAX),
to freeze the readings on the display.
again to turn off the HOLD function.
(CANCEL) to exit MIN MAX mode.
to toggle showing the T1, T2, or
.
Using the Offset to Adjust for Probe
Errors
Use the offset option in Setup to adjust the thermometer’s
readings to compensate for the errors of a specific
thermocouple.
The allowable adjustment range is
1.
2.
3.
4.
9.0
Plug the thermocouple into the input connector.
Place the thermocouple in a known, stable
temperature environment (such as an ice bath or a
dry well calibrator).
Allow the readings to stabilize.
In Setup change the offset until the primary display
reading matches the calibration temperature. (See
"Changing Setup Options.")
o
F.
5.0
o
C or K, and