FLUKE-568 Fluke Electronics, FLUKE-568 Datasheet - Page 24

THERMOMETER IR W/PROBE MENU& MEM

FLUKE-568

Manufacturer Part Number
FLUKE-568
Description
THERMOMETER IR W/PROBE MENU& MEM
Manufacturer
Fluke Electronics
Series
568r
Type
Handheld, Gunr
Datasheet

Specifications of FLUKE-568

Temperature Range
-40 ~ 1472°F (-40 ~ 800°C)
Temperature Display
C°/F°
Display Type
Dot Matrix
Input Type
Infrared; Thermocouple
Features
Alarm, Backlight, Laser Sight, Memory, MIN/MAX/Dif/Avg
Probe Type
K
Battery Type
AA (2)
Includes
Case, Software, Thermocouple - K-Type Bead Probe, USB Cable
Lead Free Status / RoHS Status
Not applicable / Not applicable
Other names
2837806
614-1073
566/568
Users Manual
16
External Contact Probe
Troubleshooting
--- (on display)
J
Blank display
Laser does not work
Failed communication via USB
Inaccuracy
Settings such as emissivity,
date/time, F/C, and saved data
lost
The thermometers come with a bead K-type thermocouple probe. The probes attach to the
thermometers via the probe input located on the top of the thermometer (see Figure 7).
With the probe installed, the probe symbol (A) appears on the display. The probe
can be used simultaneously while the thermometer is taking non-contact measurements.
The probe readings are shown below the non-contact measurements. The thermometer
will stay on when there is a probe inserted. When used with a personal computer and
FlukeView Forms (568 only), the auto off feature is disabled.
Use the following table to remedy possible problems when using the thermometer.
Symptom
To avoid electrical shock or personal injury, do not connect the
external contact probe to live electrical circuits.
Unit is off
FlukeView Forms is not running
Target temperature is over or
under range.
Low batteries
Thermometer is asleep
Possible dead batteries
Low or dead batteries
Ambient temperature is above
40 C (104 F)
Possible incorrect emissivity
setting, field of view, or spot size
Battery dead or not replaced in
<1 minute of removal
Table 4. Troubleshooting
XWWarning
Cause
Select target within
specifications
Replace Batteries
Pull trigger
Replace batteries
Replace batteries
Use in area with lower ambient
temperature
Pull the trigger
Start FlukeView Forms
See “Emissivity”, “Field of View”
and “Distance and Spot Size”
sections.
Reset settings. Replace
batteries as soon as low battery
indicated; Exchange the
batteries within one minute of
removal. Refer to “Changing the
Batteries”.
Action