FLUKE-66 Fluke Electronics, FLUKE-66 Datasheet - Page 26

THERMOMETER IR ADJ -25 TO 1100F

FLUKE-66

Manufacturer Part Number
FLUKE-66
Description
THERMOMETER IR ADJ -25 TO 1100F
Manufacturer
Fluke Electronics
Series
60r
Type
Handheld, Gunr
Datasheet

Specifications of FLUKE-66

Temperature Range
-25 ~ 1100°F (-32 ~ 600°C)
Temperature Display
C°/F°
Display Type
LCD
Input Type
Infrared; Probe
Features
Alarm, Backlight, Hold, Laser Sight, Memory, MIN/MAX/Dif/Avg
Probe Type
RTD
Battery Type
9 Volt
Includes
Case
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Other names
2148296
614-1040
63/66/68
Infrared Thermometers
CE Certification
The thermometers conform to the following standards:
Certification testing was conducted using a frequency range of 80-
1000 MHz with the instrument in three orientations.
63: Between 165 MHz and 880 MHz ( 5%) at 3 V/m, the
instrument may not meet its stated accuracy.
66/68: Between 162 MHz and 792 MHz ( 5%) at 3 V/m, the
instrument may not meet its stated accuracy.
Specifications
Temperature Range
Assumes ambient operating temperature of 23 °C (73 °F)
to 25 °C (77 °F)
63:
66:
68:
Accuracy
Target Temperature:
Above 510 °C (63)
Above 510 °C (66/68) ±1 % of reading or ±1 °C (±2 °F), whichever
23 °C to 510 °C
-18 °C to 23 °C
-26 °C to -18 °C
-32 °C to -26 °C
Resolution
63:
66/68:
22
EN61326-1 EMC
EN61010-1
EN60825-1 Safety
-32 °C to 535 °C (-25 °F to 999 °F)
-32 °C to 600 °C (-25 °F to 1100 °F)
-32 °C to 760 °C (-25 °F to 1400 °F)
±1% of reading or ±1 °C (±2 °F), whichever is
±2 °C (±3 °F)
±2.5 °C (±4 °F)
±3 °C (±5 °F)
0.2 °C (0.5 °F)
0.1 °C (0.1 °F)
is greater
greater
±1.5 % of reading
Notes