AQ0400MU24ECCFD SYNQOR, AQ0400MU24ECCFD Datasheet - Page 10

AcuQor Power Supply, Medical CF Isolation Rating

AQ0400MU24ECCFD

Manufacturer Part Number
AQ0400MU24ECCFD
Description
AcuQor Power Supply, Medical CF Isolation Rating
Manufacturer
SYNQOR
Datasheet

Specifications of AQ0400MU24ECCFD

Power Supply Type
Enclosed
Power Supply Output Type
Fixed
No. Of Outputs
1
Output Voltage
24V
Power Rating
400W
Input Voltage
85V AC To 264V AC
Length
133.4mm
Width
88.9mm
Height
41.3mm
Rohs Compliant
Yes
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Product # AQ0400MUxx Series
CONNECTOR DETAILS
OUTPUT CONNECTOR PINOUT (top side)
Pin 1 FAN_GOOD
Pin 2 AC_POWER_GOOD
Pin 3 DC_POWER_GOOD
Pin 4 5V_STANDBY
Pin 5 VOUT(+)
Pin 6 VOUT(+)
Pin 7 VOUT(+)
Pin 8 VOUT(+)
Pin 9 Reserved
Pin 10 Reserved
Pin 11 REMOTE_ENABLE
Pin 12 VOUT(–)
Pin 13 VOUT(–)
Pin 14 VOUT(–)
Pin 15 VOUT(–)
Pin 16 VOUT(–)
12 V OUTPUT CONNECTOR PINOUT (bottom side)
Pin 1 VOUT(+)
Pin 2 VOUT(+)
Pin 3 VOUT(+)
Pin 4 VOUT(+)
Pin 5 VOUT(–)
Pin 6 VOUT(–)
Pin 7 VOUT(–)
Pin 8 VOUT(–)
INDIVIDUAL INPUT CONNECTOR PINOUT
Pin 1
Pin 3
Pin 5
Technical Specification
Ground
AC Neutral
AC Line
Open collector with internal 5V pullup. See Figure A.
Pulsed low on fan failure, 100ms, 50% duty.
Short to VOUT(-) to disable fan.
Open collector with internal 5V pullup. See Figure B.
Pulled low on AC power dropout.
Open collector with internal 5V pullup. See Figure B.
Pulled low during startup ramp and within 5 ºC of
temperature shutdown threshold.
5 V @ 50 mA available whenever AC power is applied.
Positive Output Voltage.
Positive Output Voltage.
Positive Output Voltage.
Positive Output Voltage.
Reserved for future use.
Reserved for future use.
Logic input. See Figure C.
Pull high to enable main output.
Negative Output Voltage.
Negative Output Voltage.
Negative Output Voltage.
Negative Output Voltage.
Negative Output Voltage.
Positive Output Voltage.
Positive Output Voltage.
Positive Output Voltage.
Positive Output Voltage.
Negative Output Voltage.
Negative Output Voltage.
Negative Output Voltage.
Negative Output Voltage
Phone 1-888-567-9596
www.synqor.com
* Each contact rated for a maximum of 5.5 A.
Figure A: Fan status output / Fan enable input interface circuitry.
Figure B: Power good interface circuitry.
Figure C: Remote enable interface circuitry.
MATING CONNECTORS
Connector
OUTPUT (16 pins)
12V_OUTPUT (8 pins)
INPUT
AC_POWER_GOOD or
DC_POWER_GOOD
Doc.# 005-005045 Rev. B
REMOTE_ENABLE
5V_STANDBY
V
OUT
(-)
V
DC Output:
OUT
Type
Molex 430251600
Molex 436450800
JST VHR-5N
AC Input:
(-)
Power:
Grade:
0.1 µF
10 Ohms
10 Ohms
5V_STANDBY
85-264 V
12/24/36/48V
400 W Series
Medical
0.1 µF
07/22/09
5 k
5 k
N4148
N3904
5V_STANDBY
10k
Contact
Molex 430300008*
Molex 430300008*
JST SVH-41T-P1.1
RMS
From AC/DC
To Enable
Circuitry
Circuitry
Monitor
Semi-reg
Page 10
.