53210A-400 AGILENT TECHNOLOGIES, 53210A-400 Datasheet

Frequency Counter With GPIB Port, 10, 0.2 Ppm

53210A-400

Manufacturer Part Number
53210A-400
Description
Frequency Counter With GPIB Port, 10, 0.2 Ppm
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of 53210A-400

Bandwidth
350MHz
Counter Measuring Functions
Frequency, Period,Totalize, Ratio, Time Interval, Rise/Fall, Width, Burst Width, Duty Cycle, Phase, PRI, PRF
Lead Free Status / RoHS Status
na
53210A 350 MHz RF Frequency Counter, 10 digits/sec
53220A 350 MHz Universal Frequency Counter/Timer, 12 digits/sec, 100 ps
53230A 350 MHz Universal Frequency Counter/Timer, 12 digits/sec, 20 ps
Agilent 53200A Series
RF/Universal Frequency
Counter/Timers
Data Sheet

53210A-400 Summary of contents

Page 1

... Agilent 53200A Series RF/Universal Frequency Counter/Timers Data Sheet 53210A 350 MHz RF Frequency Counter, 10 digits/sec 53220A 350 MHz Universal Frequency Counter/Timer, 12 digits/sec, 100 ps 53230A 350 MHz Universal Frequency Counter/Timer, 12 digits/sec ...

Page 2

... Burst carrier frequency, pulse repetition frequency (PRF), pulse repetition interval (PRI), burst positive width (“on” time), burst negative width (“off” time). 2 Standard Opt MW Inputs 350 MHz Input (53210A Channel(s) 53220A/30A ● ● ● ● ● ...

Page 3

... Minimum signal for auto level Maximum input 50 Ω damage level 50 Ω protection threshold 1 M Ω damage level 53210A 53220A mHz) to 350 MHz (2 1000 sec 350 MHz Front panel BNC(f). Option 201 adds parallel rear panel BNC(f) inputs Selectable 1 MΩ ± 1. Ω ± 1.5% || <25 pF ...

Page 4

... Stated specification assumes Noise Reject OFF. Noise Reject ON doubles the sensitivity minimum voltage levels. 5. Assumes sine wave. 6. Assumes AM Rate > 10/gate. For Option 106, use a tolerance of 15% modulation depth for frequencies less than 900 MHz. continued 53210A 53220A 100 MHz - 6 GHz 300 MHz - 15 GHz ...

Page 5

... Acquisition length Time interval (single-shot) measurements Common Channels Single-shot time resolution Gating Slope Level Channel-to-channel time skew (typ) 53210A 53220A 10 digits/s 12 digits/s 12 digits 15 digits Reciprocal Reciprocal and resolution enhanced Continuous Wave (CW) Automatically preset or user selectable Internal or external 100 µs to 1000 µs ...

Page 6

... Range Totalize measurements Channels 9 Range Rate Gating Level measurements Voltage level - standard input channels Microwave power level (microwave channel option) continued 53210A 53220A N/A N/A N/A N/A N/A Auto-level or user selectable N/A N/A N/A N/A Auto-level or user selectable N/A .000001 to .999999 or 0.0001% to 99.9999% ...

Page 7

... Option 150 microwave pulse/burst measurement descriptions: 13. For option 115, use -10 dBm (71 mVrms) for lower sensitivity limit Detector Level Burst carrier frequency + width burst (on p-p amplitude - width burst (off) PRF PRI = 1/PRF continued 53210A 53220A 12 N/A N/A N/A N/A N/A N/A N/A N/A N/A ...

Page 8

... Connector Impedance Level Slope Gate to gate timing Damage level Gate output characteristics (typ) Connector Impedance Level Slope Damage level 53210A 53220A Time, external 100 µs - 1000 s (10 µs) N/A (unused standard channel input) N/A N/A N N/A (unused standard channel input) N/A ...

Page 9

... Damage level Timebase output (typ) Impedance Level Signal Damage level 1. Continuous, gap-free measurements limits the Gate Time setting to 10 µs to 1000 µs steps. 53210A 53220A Internal, external, bus, manual 1 to 1,000,000 3600 µs steps 1 to 1,000,000 Rear panel BNC(f) 1 kΩ ...

Page 10

... Power-off Power-on Volatile reading memory Non-volatile internal memory USB file system Capability 53210A 53220A Selectable 10 (slow), 100 (medium), 1,000 (fast) reading moving average (slow) change from average mX-b or m(1/X)-b User settable m and b (offset) values (X-b)/b scaled to %, ppm, or ppb User settable b (reference) value ...

Page 11

... LAN (VXI-11) LAN (sockets) USB GPIB Optimized (Avg. using *TRG;DATA:REM? 1, WAIT): LAN (VXI-11) LAN (sockets) USB GPIB 4 (meas) 53210A 53220A no timeout 2000 steps Slow mode (50 Hz): 350 ms (typ) Fast mode (10 kHz (typ) Frequency, Period, Range, Level (typ) 110 200 200 210 160 ...

Page 12

... REAL,64), (DISP OFF), and set fastest gate time available. 6. Maximum 53230A rates represent >= 20 MHz input signals with min gate times, no delays or holdoffs. Measurement rates for the 53210A & 53220A are limited by min gate time. Actual meas rates are limited by the repetition rate of the input being measured. 4 ...

Page 13

... Regulatory Safety EMC Acoustic noise (nom) Line power Voltage Power consumption 53210A 53220A 45-minutes 4.3" Color TFT WQVGA (480 x 272), LED backlight English, German, French, Japanese, Simplified Chinese, Korean FAT, FAT32 532xx Series and 53131A/53132A/53181A Series compatibility mode 10/ 100/ 1000 LAN (LAN Sockets and VXI-11 protocol) USB 2 ...

Page 14

... Warranty Standard 1. Assumes calibrated battery. 261.2 mm 103.8 mm 302.2 mm Dimensions apply to all three models: 53210A, 53220A, 53230A. continued 53210A 53220A Internal lithium ion battery with integrated smart battery monitor & charger °C. Battery will only charge under 35 °C. Instrument running on battery power above 50 °C will turn off to minimize battery capacity degradation. -10 ° ...

Page 15

Timebase Timebase Uncertainty = ( Aging + Temperature + Calibration Uncertainty ) Timebase Aging 1 (spec) 24-hour, T ±1 °C CAL 30-day, T ±5 °C CAL 1-year, T ±5 °C CAL Temperature (typ) 0 º ºC relative to ...

Page 16

Accuracy Specifications Definitions Random Uncertainty The RSS of all random or Type-A measurement errors expressed as the total RMS or 1-σ measurement uncertainty. Random uncertainty will reduce as 1/√N when averaging N measurement results for maximum of ...

Page 17

... For 95% Confidence use k= 2.0 in accuracy calculations. continued . Assume sharp envelope transition. IN calculation, but F E Negative width, Negative duty, Single Channel Time Interval Fall to Rise Positive width, Positive duty, Single Channel Time Interval Rise to Fall 53210A 53220A 1 use R equation E 100 ps 100 ps 100 ps 200 ps ...

Page 18

... RSS of all edge’s T alternative the cross talk from the other standard input channel. Typically this is -60 dB 53210A, and when no signal is applied to other stan- dard input channel on 53220A/53230A. (Note: the best way to eliminate cross talk is to remove the signal from the other channel). ...

Page 19

... Ordering Information Model numbers 53210A 350 MHz, 10-digits/s RF Frequency Counter 53220A 350 MHz, 12 digits/s, 100 ps Universal Frequency Counter/Timer 53230A 350 MHz, 12-digits/ Universal Frequency Counter/Timer All models include: • Certificate of Calibration and 1-year standard warranty • IEC Power Cord, USB cable • ...

Page 20

Appendix A - Worked Example Basic Accuracy Calculation for Frequency Measurement Parameter assumptions: • 53220A • 95% confidence • 100 MHz signal, 1 sec gate • AUTO frequency mode • Level input signal amplitude • TCXO standard timebase ...

Page 21

Definitions The following definitions apply to the specifications and characteristics described throughout. Specifi cation (spec) The warranted performance of a calibrated instrument that has been stored for a minimum of 2 hours within the operating temperature range of 0 °C ...

Page 22

... Spain 34 (91) 631 3300 Sweden 0200- Switzerland 0800 United Kingdom 44 (0) 118 9276201 Other European Countries: www.agilent.com/find/contactus Revised: July 8, 2010 Product specifications and descriptions in this document subject to change without notice. © Agilent Technologies, Inc. 2011 Printed in USA, March 15, 2011 5990-6283EN ...