INFS-0210-DC1 NEWPORT ELECTRONICS, INFS-0210-DC1 Datasheet - Page 49

Strain Gauge Meter

INFS-0210-DC1

Manufacturer Part Number
INFS-0210-DC1
Description
Strain Gauge Meter
Manufacturer
NEWPORT ELECTRONICS
Datasheet

Specifications of INFS-0210-DC1

No. Of Digits / Alpha
6
Meter Range
0mV To 100mV
Digit Height
13.7mm
Power Consumption
6W
Operating Temperature Range
0°C To +50°C
Signal Input Type
0 To 100mV Voltage
Lead Free Status / RoHS Status
na
8.3.2 SCALING YOUR METER WITH YOUR SENSOR CONNECTED
Before proceeding, you must first apply a low input (“INPUT 1”) and
a high input (“INPUT 2”) into your meter and record the exact
display shown. As explained in Section 8.3, the display you will be
recording will be 0 and 100000 if your meter has not been changed
from the typical factory setup and calibration. The following is an
example using the numbers below as the recorded input displayed on
the meter. You should use the numbers you have recorded:
Low input:
Meter display (“INPUT 1”):
High input:
Meter display (“INPUT 2”):
NOTE: If you are using an input signal that is reverse acting
(e.g. 100-0 mVdc), then the 100 mV reading would be “INPUT 1”
and the 0 mV reading would be “INPUT 2”.
In addition to recording your display readings for your two inputs, you
must also decide what you want these inputs to display on your meter.
In order to simplify this (especially if you want to display numbers
to the right of the decimal point position), you should think of your
meter as a process indicator and your decimal point as being passive
or cosmetic.
As an example, if you wanted your display to read 0.000 to 68.000,
you would record and enter your “READ 1” number as 000000. and
your “READ 2” number as 068000. Then after storing these
numbers, you would then place your decimal point position (refer
to Section 8.3.4) so that your display would show 0.000 to 68.000.
Record your “INPUT1” and “READ 1” numbers, and your
“INPUT2” and “READ 2” numbers, record them below and proceed
as follows:
INPUT1 =
READ 1 =
INPUT2 =
READ 2 =
8-5
0 mvdc
000000.
100 mVdc
100000.
Example
000003.
000000.
100018.
068000.
Scaling to
Display
Engineering
Units
8