DSO5012A AGILENT TECHNOLOGIES, DSO5012A Datasheet - Page 14

OSCILLOSCOPE, 100MHZ, 2 CHANNEL, 2GSPS

DSO5012A

Manufacturer Part Number
DSO5012A
Description
OSCILLOSCOPE, 100MHZ, 2 CHANNEL, 2GSPS
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of DSO5012A

Scope Type
Digital
Scope Channels
2 Analog
Bandwidth
100MHz
Meter Display Type
TFT-LCD Color
Sampling Rate
2GSPS
Input Impedance
1Mohm
Rise Time
3.5ns
Input Voltage
400V P-p
Lead Free Status / RoHS Status
na
14
Performance characteristics
Vertical system (continued)
Range
Maximum input
Offset range
Dynamic range
Input impedance
Coupling
BW limit
Channel-to-channel isolation
Standard probes
Probe ID
ESD tolerance
Noise, RMS, input shorted
DC vertical gain accuracy
DC vertical offset accuracy
Single cursor accuracy
Dual cursor accuracy
1
2
2 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 2 mV/div sensitivity setting.
Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
1
1
1
2
2 mV/div to 5 V/div (1 MΩ or 50 Ω)
Maximum input voltage for analog inputs:
CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
CAT II 100 Vrms, 400 Vpk
with N2863A 10:1 probe: CAT I 600 V, CAT II 300 V (DC + peak AC)
with 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
with 50 Ω input: 5 Vrms, CAT I
±5 V on ranges < 10 mV/div;
±20 V on ranges 10 mV/div to 200 mV/div;
±75 V on ranges >200 mV/div
±8 div
1 MΩ ± 1% || 12 pF or 50 Ω ± 1.0%, selectable
AC, DC
25 MHz selectable
DC to max bandwidth > 40 dB
DSO501xA: 10:1 N2863A shipped standard for each oscilloscope channel
DSO503xA: 10:1 N2863A shipped standard for each oscilloscope channel
DSO505xA: 10:1 10073C shipped standard for each oscilloscope channel
Auto probe sense and AutoProbe interface
Agilent- and Tektronix-compatible passive probe sense
±2 kV
DSO501xA: 0.5% full scale or 250 µV, whichever is greater
DSO503xA: 0.5% full scale or 300 µV, whichever is greater
DSO505xA: 0.5% full scale or 360 µV, whichever is greater
±2.0% full scale
≤ 200 mV/div: ±0.1 div ±2.0 mV ±0.5% offset value;
> 200 mV/div: ±0.1 div ±2.0 mV ±1.5% offset value
±{DC vertical gain accuracy + DC vertical offset accuracy + 0.2% full scale (~1/2 LSB)}
Example: for 50 mV signal, oscilloscope set to 10 mV/div (80 mV full scale), 5 mV offset,
accuracy = ±{2.0% (80 mV) + 0.1 (10 mV) + 2.0 mV + 0.5% (5 mV) + 0.2%(80 mV)} = ± 4.785 mV
±{DC vertical gain accuracy + 0.4% full scale (~1 LSB)}
Example: for 50 mV signal, oscilloscope set to 10 mV/div (80 mV full scale), 5 mV offset,
accuracy = ±{2.0% (80 mV) + 0.4% (80 mV)} = ±1.92 mV
(continued)