DSO5012A AGILENT TECHNOLOGIES, DSO5012A Datasheet - Page 19

OSCILLOSCOPE, 100MHZ, 2 CHANNEL, 2GSPS

DSO5012A

Manufacturer Part Number
DSO5012A
Description
OSCILLOSCOPE, 100MHZ, 2 CHANNEL, 2GSPS
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of DSO5012A

Scope Type
Digital
Scope Channels
2 Analog
Bandwidth
100MHz
Meter Display Type
TFT-LCD Color
Sampling Rate
2GSPS
Input Impedance
1Mohm
Rise Time
3.5ns
Input Voltage
400V P-p
Lead Free Status / RoHS Status
na
Performance characteristics
General characteristics
Physical size
Weight
Probe comp output frequency ~
Trigger out
Kensington lock
Power requirements
Line voltage range
Line frequency
Power usage
Environmental characteristics
Ambient temperature
Humidity operating
Altitude
Vibration
Shock
Pollution degree
Indoor use
Other
Measurement categories
Regulatory information
Supplementary information
35.4 cm wide x 18.8 cm high x 17.4 cm deep (without handle)
38.5 cm wide x 18.8 cm high x 17.4 cm deep (with handle)
Net: 4.1 kg (9 lbs)
Shipping: approximately 9 kgs (20 lbs)
1.2 kHz, amplitude ~2.5 V
0 to 5 V into open circuit (~23 ns delay)
0 to 2.5 V into 50 Ω
Connection on rear panel for security
100-120 V, 50/60/400 Hz; 100-240V, 50/60 Hz auto ranging
50/60 Hz, 100-240 VAC; 440 Hz, 100-132 VAC
120 W max
Operating -10 °C to +55 °C; non-operating –40 °C to +70 °C
95% RH at 40 °C for 24 hour; non-operating 90% RH at 65 °C for 24 hour
Operating to 4,570 m (15,000 ft); non-operating to 15,244 m (50,000 ft)
Agilent class GP and MIL-PRF-28800F; Class 3 random
Agilent class GP and MIL-PRF-28800F (operating 30 g, 1/2 sine, 11-ms duration, 3 shocks/axis along
major axis. Total of 18 shocks)
Normally only dry non-conductive pollution occurs. Occasionally a temporary conductivity caused by
condensation must be expected.
Rated for indoor use only
CAT I
Safety IEC 61010-1:2001 / EN 61010-1:2001
Canada: CSA C22.2 No. 61010-1:2004
USA: UL 61010-1:2004
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and the
EMC Directive 89/336/EEC, and carries the CE-marking accordingly.
The product was tested in a typical configuration with HP/Agilent test systems.
Product specifications, characteristics, and descriptions in this document are subject to change
without notice.
(continued)
19