MSO4104B Tektronix, MSO4104B Datasheet - Page 15

no-image

MSO4104B

Manufacturer Part Number
MSO4104B
Description
OSCILLOSCOPE, 20 CHANNEL, 1GHZ, 5GSPS
Manufacturer
Tektronix
Type
Mixed Signalr
Datasheet

Specifications of MSO4104B

Scope Type
Digital Bench
Scope Channels
4 Analog + 16 Digital
Bandwidth
1GHz
Meter Display Type
XGA LCD
Sampling Rate
5GSPS
Input Impedance
1Mohm
Rise Time
350ps
Input Voltage
250Vrms
Configuration
Benchtop
Number Of Channels
4 Analog + 16 Digital
Sample Rate
5 GS/s
Rohs Compliant
NA
Lead Free Status / RoHS Status
na
Limit/Mask Testing (Optional)
Software
Characteristic
Included Standard Masks ITU-T, ANSI T1.102, USB
Test Source
Mask Creation
Mask Scaling
Test Criteria Run Until
Violation Threshold
Actions on Test Failure
Actions on Test Complete Trigger out pulse, set remote interface SRQ
Results Display
Software
NI LabVIEW
SignalExpress Tektronix
Edition
OpenChoice
IVI Driver
e*Scope
LXI Class C
Web Interface
®
®
Desktop
Description
Limit Test: Any Ch1 - Ch4 or any R1 - R4
Mask Test: Any Ch1 - Ch4
Limit test vertical tolerance from 0 to 1 division in 1 m
division increments; Limit test horizontal tolerance from
0 to 500 m division in 1 m division increments
Load standard mask from internal memory
Load custom mask from text file with up to 8 segments
Lock to Source ON (mask automatically re-scales with
source-channel settings changes)
Lock to Source OFF (mask does not re-scale with
source-channel settings changes)
Minimum number of waveforms (from 1 to 1,000,000;
Infinity)
Minimum elapsed time (from 1 second to 48 hours;
Infinity)
From 1 to 1,000,000
Stop acquisition, save screen image to file, save
waveform to file, print screen image, trigger out pulse,
set remote interface SRQ
Test status, total waveforms, number of violations,
violation rate, total tests, failed tests, test failure rate,
elapsed time, total hits for each mask segment
Description
A fully interactive measurement software environment
optimized for the MSO/DPO4000B Series, enables
you to instantly acquire, generate, analyze, compare,
import, and save measurement data and signals using
an intuitive drag-and-drop user interface that does not
require any programming.
Standard MSO/DPO4000B Series support for
acquiring, controlling, viewing, and exporting your
live signal data is permanently available through
the software. The full version (SIGEXPTE) adds
additional signal processing, advanced analysis, mixed
signal, sweeping, limit testing, and user-defined step
capabilities and is available for a 30-day trial period
standard with each instrument.
Enables fast and easy communication between a
Windows PC and the MSO/DPO4000B Series, using
USB or LAN. Transfer and save settings, waveforms,
measurements, and screen images. Included Word
and Excel toolbars automate the transfer of acquisition
data and screen images from the oscilloscope into
Word and Excel for quick reporting or further analysis.
Provides a standard instrument programming
interface for common applications such as LabVIEW,
LabWindows/CVI, Microsoft .NET and MATLAB.
Enables control of the MSO/DPO4000B Series over a
network connection through a standard web browser.
Simply enter the IP address or network name of the
oscilloscope followed by :81 (to represent connecting
through Port 81) and a web page will be served to the
browser
Connect to the MSO/DPO4000B Series through
a standard web browser by simply entering the
oscilloscope’s IP address or network name in the
address bar of the browser. The web interface enables
viewing of instrument status and configuration, and
status and modification of network settings. All web
interaction conforms to LXI Class C specification.
Mixed Signal Oscilloscopes — MSO4000B, DPO4000B Series
Display Characteristics
Input/Output Ports
LAN eXtensions for Instrumentation (LXI)
Power Source
Characteristic
Display Type
Display Resolution
Waveform Styles
Graticules
Format
Waveform Capture Rate
Port
USB 2.0 High-speed Host
Port
USB 2.0 Device Port
LAN Port
XGA Video Port
Auxiliary Input
Probe Compensator
Output
Auxiliary Out
External Reference In
Kensington Lock
Characteristic
Class
Version
Characteristic
Power Source Voltage
Power Source Frequency 45 to 66 Hz (85 to 264 V)
Power Consumption
Description
Description
10.4 in. (264 mm) liquid-crystal TFT color display
1,024 horizontal × 768 vertical pixels (XGA)
Vectors, Dots, Variable Persistence, Infinite
Persistence
Full, Grid, Solid, Cross Hair, Frame, IRE, and mV
YT and simultaneous XY/YT
>50,000 wfm/s maximum
Supports USB mass storage devices, printers and
keyboard. Two ports on front and two ports on rear of
instrument
Rear-panel connector allows for communication/control
of oscilloscope through USBTMC or GPIB
(with a TEK-USB-488), and direct printing to all
PictBridge-compatible printers
RJ-45 connector, supports 10/100/1000 Mb/s
DB-15 female connector, connect to show the
oscilloscope display on an external monitor or projector
Front-panel BNC connector. Input Impedance 1 MΩ.
Max input 300 V
Front-panel pins
Amplitude: 2.5 V
Frequency: 1 kHz
Rear-panel BNC connector
V
V
to ground
Output can be configured to provide a pulse out signal
when the oscilloscope triggers, the internal oscilloscope
reference clock out, or an event out for limit/mask
testing
Time-base system can phase lock to an external
10 MHz reference (10 MHz ±1%)
Rear-panel security slot connects to standard
Kensington lock
Description
LXI Class C
V1.3
Description
100 to 240 V ±10%
360 to 440 Hz (100 to 132 V)
225 W maximum
OUT
OUT
(Hi): ≥2.5 V open circuit, ≥1.0 V 50 Ω to ground
(Lo): ≤0.7 V into a load of ≤4 mA; ≤0.25 V 50 Ω
RMS
CAT II with peaks ≤ ±425 V
www.tektronix.com
15