81150A-002 AGILENT TECHNOLOGIES, 81150A-002 Datasheet - Page 30

NOISE GENERATOR, ARB/FREQ/PULSE, 120MHZ

81150A-002

Manufacturer Part Number
81150A-002
Description
NOISE GENERATOR, ARB/FREQ/PULSE, 120MHZ
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of 81150A-002

Signal Generator Type
Pulse
Bandwidth
120MHz
Modulation Type
Amplitude/Frequency/Phase/Pulse
Supply Voltage Range
100V To 240V
External Height
108mm
External Width
439mm
Lead Free Status / RoHS Status
na
General Specifi cations
Defi nitions
Specifi cation (spec.)
The warranted performance of a calibrated instrument that has been stored for a minimum of 2 hours within the operating temperature
range of 0 °C to 55 °C and after a 45-minute warm up period. Within ± 20 °C after autocal. All specifi cations include measurement
uncertainty and were created in compliance with ISO-17025 methods. Data published in this document are specifi cations (spec) only
where specifi cally indicated.
Typical (typ.)
The characteristic performance, which 80% or more of manufactured instruments will meet. This data is not warranted, does not include
measurement uncertainty, and is valid only at room temperature (approximately 23 °C).
Nominal (nom.)
The mean or average characteristic performance, or the value of an attribute that is determined by design such as a connector type,
physical dimension, or operating speed. This data is not warranted and is measured at room temperature (approximately 23 °C).
Measured (meas.)
An attribute measured during development for purposes of communicating the expected performance. This data is not warranted and is
measured at room temperature (approximately 23 °C).
Accuracy
Represents the traceable accuracy of a specifi ed parameter. Includes measurement error and timebase error, and calibration source
uncertainty.
Power supply
Power consumption
Operating temperature
Operating altitude
Storage temp.
Stored states
Power on state
Interface
Programming language
Dimensions (WxHxD)
Bench top
Rack mount
Weight
Safety designed to
EMC tested to
Warm up time
Calibration interval
Warranty
Cooling requirements
81150A
100 V to 240 V ~, 50 to 60 Hz
100 V to 127 V ~, 50 to 400 Hz
180 W max.
0 to 50 ºC
Up to 2000 m
-40 to 70 °C
4 named user confi gurations and factory default
Default or last state
2 x USB 2.0 standard A,
1 x USB 2.0 standard B,
GPIB and LAN
SCPI-1997
IEEE-488.2
LXI compliant to LXI class C
439 mm x 108 mm x 456 mm
428 mm x 89 mm x 439 mm
8 kg
IEC61010-1
UL61010
CSA22.2 61010.1 certifi ed
IEC61326
30 min.
1 year recommended
1 year standard
When operating the instrument choose a location that
provides at least 80 mm of clearance at rear, and at
least 30mm of clearance at each side
30
81160A
100 V to 240 V ~, 50 to 60 Hz
100 V to 127 V ~, 50 to 400 Hz
90 W nom.
0 to 55 ºC
Up to 2000 m
-40 to 70 °C
4 named user confi gurations and factory default
Default or last state
2 x USB 2.0 standard A,
1 x USB 2.0 standard B,
GPIB and LAN
SCPI-1997
IEEE-488.2
LXI compliant to LXI class C
439 mm x 108 mm x 456 mm
428 mm x 89 mm x 439 mm
8 kg
IEC61010-1
UL61010
CSA22.2 61010.1 certifi ed
IEC61326
30 min.
2 years recommended
1 year standard
When operating the instrument choose a location that
provides at least 80 mm of clearance at rear, and at