8808A 120V Fluke Electronics, 8808A 120V Datasheet - Page 13

DMM 5.5 RES RS-232 INTERFACE

8808A 120V

Manufacturer Part Number
8808A 120V
Description
DMM 5.5 RES RS-232 INTERFACE
Manufacturer
Fluke Electronics
Type
Digital (DMM)r
Datasheets

Specifications of 8808A 120V

Includes
Test Leads
Style
Bench
Display Digits
5.5
Display Type
VFD, Dual
Display Count
200000
Function
Voltage, Current, Resistance, Frequency
Functions, Extra
Continuity, dB, Diode Test
Features
Hold, Min/Max, RS-232 Port
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Not applicable / RoHS non-compliant
Other names
614-1068
To replace the current-input fuses:
To avoid electric shock, remove the power cord and any test
leads from the Meter before opening the current-input fuse
access door.
1. Unplug the power cord.
2. Turn the Meter upside down.
3. Remove the retaining screw on the fuse access door. See Figure 2.
4. Remove the protective cover from the fuse holders by slightly depressing the
5. Remove the defective fuse and replace it with a fuse of an appropriate rating. See
6. Replace the protective cover by pushing it over the fuses while aligning the
7. Replace the fuse access door and install the retaining screw.
back edge of the cover to unlatch it from the printed circuit board. Pull up on the
back edge of the cover and remove it from the fuse compartment.
Table 3.
catches with the holes in the printed circuit board. Press the cover down until the
catches engage the printed circuit board.
Figure 2. Replacing the Current-Input Fuses
Fuses
F1
F2
XW Warning
Bottom front
left corner
Power Considerations
Digital Multimeter
caw20.eps
7