BK2534 B&K Precision, BK2534 Datasheet - Page 25

OSCOPE 60MHZ DIGITAL STORAGE

BK2534

Manufacturer Part Number
BK2534
Description
OSCOPE 60MHZ DIGITAL STORAGE
Manufacturer
B&K Precision
Type
Bench, Digitalr
Datasheets

Specifications of BK2534

Includes
Probes, Software, USB Cable
Bandwidth
60MHz
Channels
Dual
Display Type
LCD - Color
Features
Software Controllable, USB Port, Waveform Capture
Probe Type
PR-37A x1/x10/Ref Probes (2)
Sampling Rate (per Second)
400MSa/s
Input Impedance
1M - 19pF
Rise Time (typ)
5.83nS
Voltage - Input, Max
400Vpeak
Voltage - Supply
100 ~ 240VAC
Lead Free Status / RoHS Status
Not applicable / Not applicable
X-Y Horizontal Mode
X-Y mode changes the display from a volts-versus-time
display to a volts-versus-volts display. The time base is
turned off. The CH1 signal is plotted on the X axis and CH2
signal is plotted on the Y axis.
You can use X-Y mode to compare the frequency and phase
relationships between two signals. X-Y mode can also be
used with voltage transducers to display, for example, strain
versus displacement, flow versus pressure, volts versus
current, or voltage versus frequency.
In order to get a better view of the waveform, the proper
vertical scale should be selected before entering the X-Y
mode.
Example: Use the X-Y mode to compare two signals with
the
same frequency and different phase. Connect the two signals
to CH1 and CH2 respectively. Press horizontal MENU key
and then X-Y softkey to select X-Y mode. The displayed
ellipse (a Lissajous figure) indicates the phase relationship of
the two signals.
 
Basic Operation
 
49
50
Basic Operation
 
 
X-Y Horizontal Mode