Defi nitions and Conditions
Specifications describe the performance of parameters covered by the product
warranty and apply to the full temperature of 0 to 55 °C
noted.
95th percentile values indicate the breadth of the population (approx. 2 σ) of
performance tolerances expected to be met in 95 percent of the cases with a 95
percent confidence, for any ambient temperature in the range of 20 to 30 °C. In
addition to the statistical observations of a sample of instruments, these values
include the effects of the uncertainties of external calibration references. These
values are not warranted. These values are updated occasionally if a significant
change in the statistically observed behavior of production instruments is
observed.
Typical describes additional product performance information that is not covered
by the product warranty. It is performance beyond specifications that 80 percent
of the units exhibit with a 95 percent confidence level over the temperature
range 20 to 30 °C. Typical performance does not include measurement uncer-
tainty.
Nominal values indicate expected performance, or describe product performance
that is useful in the application of the product, but are not covered by the prod-
uct warranty.
The analyzer will meet its specifications when:
• It is within its calibration cycle
• Under auto couple control, except when Auto Sweep Time Rules = Accy
• Signal frequencies < 10 MHz, with DC coupling applied
• The analyzer has been stored at an ambient temperature within the allowed
• The analyzer has been turned on at least 30 minutes with Auto Align set to
1. For earlier instruments (Serial number prefix < MY/SG/US5052), the full temperature ranges from
operating range for at least two hours before being turned on, if it had previ-
ously been stored at a temperature range inside the allowed storage range,
but outside the allowed operating range
normal, or, if Auto Align is set to off or partial, alignments must have been run
recently enough to prevent an Alert message; if the Alert condition is changed
from Time and Temperature to one of the disabled duration choices, the
analyzer may fail to meet specifications without informing the user
5 to 50 °C.
1
, unless otherwise
3
This EXA signal analyzer data
sheet is a summary of the
complete specifications and
conditions for N9010A EXA signal
analyzers (including N9010AEP
Express EXA signal analyzers),
which are available in the EXA
Signal Analyzer Specification
Guide. The EXA Signal Analyzer
Specification Guide can be
obtained on the web at:
www.agilent.com/find/exa_manuals
For ordering information, refer
to the EXA Signal Analyzer
Configuration Guide
(5989-6531EN).