72-8155 TENMA, 72-8155 Datasheet - Page 12

LCR Meter With HFE Transistor

72-8155

Manufacturer Part Number
72-8155
Description
LCR Meter With HFE Transistor
Manufacturer
TENMA
Datasheet

Specifications of 72-8155

Accuracy %
4%
External Width
3.27"
External Depth
1.5"
No. Of Batteries
1
Kit Contents
LCR Meter, Test Clips, Owners Manual, 1 9V Battery
Capacitance Measuring Range
2nF To 600µF
Lead Free Status / RoHS Status
na
Note
l
l
l
l
When measuring at 20
can add 0.1 to 0.3 error to resistance. To obtain precise
readings in these low-resistance measurement, that is
the range 20
beforehand and record the reading obtained (called this
reading as X). (X) is the additional resistance from the
test clips.
Then use the equation:
measured resistance value (Y) – (X) = precision readings
of resistance.
The Meter displays “1” when there is no input, for
example, open circuit situation.
For high resistance measurement (>1M ), it is normal
to take several seconds to obtain a stable reading.
When resistance measurement has been completed,
disconnect the connection between the test clips and
the circuit under test and remove the test clips from the
input terminals of the Meter.
and 200 , short circuit the input terminals
and 200
12
Model 72-8155: OPERATING MANUAL
range, the test clips