U2353A AGILENT TECHNOLOGIES, U2353A Datasheet - Page 51

USB Modular Data Acquisition

U2353A

Manufacturer Part Number
U2353A
Description
USB Modular Data Acquisition
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of U2353A

No. Of Channels
16
Sampling Rate
1MSPS
Input Voltage Max
12V
Input Current Max
2A
Bandwidth
400kHz
External Height
44mm
External Width
120mm
External Depth
182.4mm
Sample Rate
500 KSa/s
Lead Free Status / RoHS Status
na
Performance characteristics
[1]
Vertical system: oscilloscope channels (continued)
Single-cursor accuracy
Dual-cursor accuracy
Horizontal
Range
Time base accuracy
Delay range
Modes
XY
Reference position
Trigger System
Sources
Modes
Holdoff time
Selections
• Edge
• Pulse width
• TV
Autoscale
Oscilloscope channel triggering
Range (internal)
Coupling
External (EXT) triggering
Input impedance
Maximum input
Range
EXT trigger pulse width
All specifi cations are warranted. specifi cations are valid after a 30-minute warm-up and within ±100 ºC of last calibration temperature.
[1]
(continued)
±{DC vertical gain accuracy + DC vertical offset accuracy + 0.2% full scale (~½ LSB)}
Example:
For 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,
Accuracy = ±{4.0% (80 mV) + 0.1(10 mV) + 2.0 mV + 0.5% (5 mV) + 0.2% (80 mV)}
±{DC vertical gain accuracy + 0.4% full scale (~1 LSB)}
Example:
For 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,
Accuracy = ±{4.0% (80 mV) + 0.4% (80 mV)}
1 ns/div to 50 s/div
20 ppm
Pre-trigger: –100 %
Post-trigger: +100 %
Main, roll, XY
Yes
Center
Ch 1, Ch 2, Ext (not applicable for TV trigger)
Normal, single, auto trigger
60 ns
Edge, pulse width, TV
Triggers on a rising or falling edge, alternating, or either edge of any source
Triggers on a pulse width greater than, equal to, or less than a specifi ed time limit, with
time limits ranging from 16 ns to 10 s.
• Minimum lower limit: 8 ns
• Minimum upper limit: 16 ns
• Maximum pulse width setting: 10 s
Triggers on one of three standard television waveforms: NTSC, PAL, SECAM
TV trigger sensitivity: 0.6 division of sync signal. Modes supported include Field 1, Field 2,
all fi elds, or any line within a fi eld.
Single-button automatic setup of all channels
±4 div from center screen
AC (< 15 Hz)
LF reject (~ 35 kHz)
HF reject (~ 35 kHz)
1 MW: ≈ 16 pF
CAT I 30 Vrms, 42 Vpk
DC coupling: trigger level ±1.25 V and ±2.5 V
> 2.5 ns
= ±6.385 mV
= ±3.52 mV