FLUKE-772 Fluke, FLUKE-772 Datasheet - Page 9

MULTIMETER, DIGITAL, CLAMP

FLUKE-772

Manufacturer Part Number
FLUKE-772
Description
MULTIMETER, DIGITAL, CLAMP
Manufacturer
Fluke
Datasheets

Specifications of FLUKE-772

Dmm Type
Clamp
Voltage Measuring Range Dc
0 To 30.00 V Dc
Current Measuring Range Dc
0mA To 100mA
Current Measuring Range Ac
0mA To 100mA
Dmm Functions
AC/DC Current
Current Ac Max
100mA
Ranging
Auto
Altitude
0 to 2000 m
Dimensions
1.63" × 3.0" × 9.75"
Humidity
<90% @ <30°C, <75% @ 30 to 55°C
Ip Rating
IP 40
Shock Resistance
1 m Drop Test (except jaw)
Temperature, Operating
-10 to 50 °C
Vibration
Random 2 g, 5 to 500 Hz
Lead Free Status / RoHS Status
na
Lead Free Status / RoHS Status
na
Measurements can be taken with the clamp in the docked position,
remotely using the 1 m cable, or via test leads. For accurate
measurements:
To use the clamp for measurements:
1.
2.
To avoid electric shock, do not use the clamp on non-
insulated conductors.
Always zero the Meter prior to taking measurements with the clamp.
To reduce magnetic influences, zero the Meter as close to the
measurement in the same position or jaw direction that is used for
the measurements as possible.
Make sure the clamp is free of contamination.
Press C to enter Clamp Measure mode and to zero the Meter.
Clamp mode includes clamp measure, mA scale output, and mA
IN/OUT. If necessary, press N to change to mA scale.
Clamp the jaw around the conductor under test. The Meter displays
the measured conductor current. See Figure 6.
The small secondary display shows the reading in mA percentage of
span.
A positive reading indicates current flowing in the direction of
the arrow on the clamp.
A negative reading indicates current flowing in the opposite
direction of the arrow.
Do not clamp more than one wire.
Figure 6. Taking Measurements with the Clamp
XW Warning
OK
fjv03.eps