1132A AGILENT TECHNOLOGIES, 1132A Datasheet - Page 28

TEST PROBE, OSCILLOSCOPE

1132A

Manufacturer Part Number
1132A
Description
TEST PROBE, OSCILLOSCOPE
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of 1132A

Test Probe Type
Oscilloscope
Test Probe Functions
High Frequency
Accuracy
± 3%
For Use With
Infiniium Series Oscilliscopes
Lead Free Status / RoHS Status
na
Current Probes Overview
Current probes sense the current
flowing through a conductor and
convert it to a voltage that can be
viewed and measured on an oscil-
loscope. There are two types of
current probes for oscilloscopes,
ac current probes (usually pas-
sive probes) and ac/dc current
probes (usually active probes).
Both types use the same principle
of transformer action for sensing
ac current in a conductor. The
alternating current flowing
through a conductor causes a flux
field to build and collapse accord-
ing to the amplitude and direc-
tion of current flow. The ac cur-
rent probe head is actually a coil—
when placed in this field the
changing flux field induces a volt-
age across the coil. The ac/dc cur-
rent probes also contain a Hall
Effect device to sense dc and low
frequency ac. Because dc doesn’t
cause a changing flux field it can-
not be sensed by transformer
action. Also, at frequencies very
close to dc, the flux field may not
change fast enough for measura-
ble transformer action.
Model
1146A
1147A
Current Probe Advantages
1146A low-cost model measures ac and dc current
to 100 Arms without breaking into the circuit
1147A measures ac and dc current to 50 MHz
without breaking into the circuit
28
Probe Type
ac/dc Current
ac/dc Current
Figure 5.1. Agilent 1146A low-cost
ac/dc current probe.
Applications and Use
Measure dc and ac current simultaneously. Useful for
lines type power measurements.
Measure dc and ac current simultaneously. Useful for
pwm/switching power measurements.
Limitations
100 kHz bandwidth
15 A rated current
Figure 5.2. Agilent 1147A general-
purpose ac/dc current probe.
Page
29
31