N9010A-503 Agilent Technologies Test Equipment, N9010A-503 Datasheet - Page 10

no-image

N9010A-503

Manufacturer Part Number
N9010A-503
Description
EXA SIGNAL ANALYZER, 3 GHZ
Manufacturer
Agilent Technologies Test Equipment
Type
Networkr
Datasheet

Specifications of N9010A-503

Frequency
3 GHz
Lead Free Status / Rohs Status
RoHS Exempt Product
Dynamic Range Specifications
1 dB gain compression (two-tone)
Preamp on (Option P03)
Displayed average noise level (DANL)
(Input terminated, sample or average detector, averaging type = Log, 0 dB input attenuation, IF Gain = High, 20 to 30 °C)
Preamp off
Preamp on (Option P03)
Spurious responses
Residual responses (Input
terminated and 0 dB attenuation)
Image responses
LO related spurious
(f > 600 MHz from carrier)
Other spurious
10
First RF order
Higher RF order
f ≥ 10 MHz from carrier
f ≥ 10 MHz from carrier
20 MHz to 26.5 GHz
10 MHz to 3.6 GHz
1 to 10 MHz
10 MHz to 2.1 GHz
2.1 to 3.6 GHz
3.6 to 7.0 GHz
7.0 to 13.6 GHz
13.6 to 17.1 GHz
17.1 to 20.0 GHz
20.0 to 26.5 GHz
10 MHz to 2.1 GHz
2.1 to 3.6 GHz
200 kHz to 8.4 GHz (swept)
Zero span or FFT or other frequencies –100 dBm nominal
10 MHz to 3.6 GHz
3.6 to 13.6 GHz
13.6 to 17.1 GHz
17.1 to 22 GHz
22 to 26.5 GHz
10 MHz to 3.6 GHz
–68 dBc
–80 dBc
Total power at input mixer
+9 dBm nominal
–10 dBm nominal
Specifi cation
–147 dBm
–148 dBm
–147 dBm
–147 dBm
–143 dBm
–137 dBm
–137 dBm
–134 dBm
–161 dBm
–160 dBm
–100 dBm
–80 dBc (–103 dBc typical)
–75 dBc (–87 dBc typical)
–71 dBc (–85 dBc typical)
–68 dBc (–82 dBc typical)
–66 dBc (–78 dBc typical)
–90 dBc typical
Typical
–149 dBm
–150 dBm
–148 dBm
–149 dBm
–147 dBm
–142 dBm
–142 dBm
–140 dBm
–163 dBm
–162 dBm