N9010A-503 Agilent Technologies Test Equipment, N9010A-503 Datasheet - Page 5

no-image

N9010A-503

Manufacturer Part Number
N9010A-503
Description
EXA SIGNAL ANALYZER, 3 GHZ
Manufacturer
Agilent Technologies Test Equipment
Type
Networkr
Datasheet

Specifications of N9010A-503

Frequency
3 GHz
Lead Free Status / Rohs Status
RoHS Exempt Product
Frequency and Time Specifications
Frequency span (FFT and swept mode)
Range
Resolution
Accuracy
Sweep time and triggering
Range
Accuracy
Trigger
Trigger delay
Time gating
Gate methods:
Gate length range (except method = FFT):
Gate delay range:
Gate delay jitter:
Sweep (trace) point range
All spans
Resolution bandwidth (RBW)
Range (–3.01 dB bandwidth)
Bandwidth accuracy (power)
RBW range
Bandwidth accuracy (–3.01 dB)
RBW range
Selectivity (–60 dB/–3 dB)
Swept
FFT
0 Hz (zero span), 10 Hz to maximum frequency of instrument
2 Hz
±(0.25% x span + horizontal resolution)
±(0.10% x span + horizontal resolution)
Span = 0 Hz
Span ≥ 10 Hz
Span ≥ 10 Hz, swept
Span ≥ 10 Hz, FFT
Span = 0 Hz
Free run, line, video, external 1, external 2, RF burst, periodic timer
Span = 0 Hz or FFT
Span ≥ 10 Hz, swept
Resolution
Gated LO; Gated video; Gated FFT
100.0 ns to 5.0 s
0 to 100.0 s
33.3 ns p-p nominal
1 to 40001
1 Hz to 3 MHz (10% steps), 4, 5, 6, 8 MHz
1 Hz to 750 kHz
820 kHz to 1.2 MHz (< 3.6 GHz CF)
1.3 to 2.0 MHz (< 3.6 GHz CF)
2.2 to 3 MHz (< 3.6 GHz CF)
4 to 8 MHz (< 3.6 GHz CF)
1 Hz to 1.3 MHz
4.1:1 nominal
(continued)
1 µs to 6000 s
1 ms to 4000 s
±0.01% nominal
±40% nominal
±0.01% nominal
–150 to +500 ms
1 µs to 500 ms
0.1 µs
±1.0% (±0.044 dB)
±2.0% (±0.088 dB)
±0.07 dB nominal
±0.15 dB nominal
±0.25 dB nominal
±2% nominal
5