LU27-01 Flowline, LU27-01 Datasheet

no-image

LU27-01

Manufacturer Part Number
LU27-01
Description
Sensor, Ultrasonic, 8.2 ft Range, 1 in NPT, NEMA 6P, 10 ft Cable, cFMus w/Fob
Manufacturer
Flowline
Type
Level Transmitterr
Datasheet

Specifications of LU27-01

Body Style
Compact
Brand/series
EchoSonic
Function
Non-contact Level Measurement
Ip Rating
IP67
Mounting Type
1 In. NPT
Output
4-20 mA
Range, Measurement
8.2 Ft.
Sensing Mode
Ultrasonic
Special Features
Unmatched Beam Width
Termination
Cable
Voltage, Supply
24 VDC
For Use With
Level Measurement
EchoSonic II
Description
The general purpose two-wire ultrasonic
transmitter provides non-contact level
measurement up to 32.8' (10 m), and is
well suited for challenging ultrapure, cor-
rosive, slurry or waste liquids. The trans-
mitter is broadly selected for atmospheric
bulk storage, day tank, waste sump, pro-
cess vessel, IBC tote, 55-gallon drum
and waste sump applications. Media
examples include acetic acid and resin.
Specifications
Range:
Accuracy:
Resolution:
Beam width:
Dead band:
Memory:
Supply voltage:
Consumption:
Loop resist.:
Signal output:
Signal invert:
Configuration:
Fail-safety:
Process temp.:
Temp. comp.:
Ambient temp.:
Pressure:
Enclosure rating:
Encl. material:
Trans. material:
Cable jacket mat.:
Cable type:
Cable length:
Process mount:
Mount. gasket:
Classification:
Compliance:
Approvals:
LU27: 4" to 8.2'
(10 cm to 2.5 m)
LU28: 8" to 24.6'
(20 cm to 7.5 m)
LU29: 8" to 32.8'
(20 cm to 10 m)
± 0.2% Range
LU27: 0.019” (0.5mm)
LU28/29: 0.079" (2 mm)
LU27: 2" (5 cm) dia.
LU28/29: 3" (7.6 cm) dia.
LU27: 4" (10 cm)
LU28/29: 8" (20 cm)
Non-volatile
24 VDC
0.5W
500Ω Ohms @ 24 VDC
4-20 mA, two-wire
4-20 mA or 20-4 mA
WebCal® PC Windows®
USB 2.0
4 mA, 20 mA, 21 mA,
22 mA or hold last
F: -4˚ to 140˚
C: -20˚ to 60˚
Automatic
F: -31˚ to 140˚
C: -35˚ to 60˚
MWP = 30 PSI
Type 6P, encapsulated,
corrosion resistant &
submersible
Polycarbonate
PVDF
Polyurethane
4-cond., shielded
10' (3m)
LU27: 1" NPT (1" G)
LU28/29: 2" NPT (2" G)
Viton®
General purpose
RoHS
cFMus, CE
EchoSonic II
Sensor range / size
Process mount
27
28
29
0
1
Features
WebCal, an innovative PC user inter-
face that provides fast and accurate
configuration
Offered in 3 measurement ranges up to
32.8' (10 m)
Optimize
unmatched beam width for applications
with restricted space
Industry leading miniature dead band maxi-
mizes the filling capacity of vessels
Polycarbonate enclosure rated NEMA
6P with PVDF transducer
2.5 m (8.2')
7.5 m (24.6')
10 m (32.8')
NPT (US)
G (Metric)
LU28-00
Shown
LU
sensor
Ordering
placement
TM
-
Non-Contact Ultrasonic Level Transmitter
with
1
Ordering Notes
1) For installation on open tanks, order the
2) WebCal is a free download from our website
Configuration
EchoPod is configured using Flowline’s
WebCal software and Fob, a USB® in-
terface tool. A single configuration can
be easily created, saved, duplicated,
emailed or modified. Take control of
your level process with WebCal’s intui-
tive interface, pre-programmed menus,
tank set point graphics and custom
wiring diagrams for each of your con-
figurations. It’s level configuration made
simple.
LM50-1001-1 (1" NPT) or LM50-1061-1 (1" G)
s i d e m o u n t b r a c k e t ; o r o r d e r t h e
LM50-1001 (2" NPT) or LM50-1061 (2" G)
side mount bracket.
at www.flowline.com/webcal (Windows® XP
Compatible, USB 2.0).
WebCal Software
LI99-1001
Shown

Related parts for LU27-01

Related keywords