4
66
CAU T ION
Performance Test
AC Voltage Verification Test
Configuration: AC volts
1 Select the AC voltage function.
2 Select each range in the order shown in
Table 4-7 AC volts verification test
Set the calibrator output to 0 V before disconnecting it from the multimeter input
terminals.
Function
AC voltage
indicated input voltage and frequency. Compare measurement results to
the appropriate test limits shown in
appropriate source settling).
V RMS
50.00 mV
450.00 mV
0.5000 V
4.5000 V
5.000 V
45.000 V
50.00 V
450.00 V
75.0 V
675.0 V
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
Input frequency
Table 4-
Range
500 mV
500 mV
5 V
5 V
50 V
50 V
500 V
500 V
750 V
750 V
Table 4-
U3401A User’s and Service Guide
7. (Be certain to allow for
7. Provide the
Error from nomial
one year
±650 μV
±2.65 mV
±3.2 mV
±17.2 mV
±32 mV
±172 mV
±400 mV
±2.4 V
±1.8 V
±4.8 V