CAU T ION
Performance Test
AC Voltage Verification Test
Configuration: AC volts
1 Select the AC voltage function.
2 Select each range in the order shown in
Table 4-8 AC volts verification test
Set the calibrator output to 0 V before disconnecting it from the multimeter
input terminals.
Function
AC voltage
input voltage and frequency. Compare measurement results to the
appropriate test limits shown in
appropriate source settling.)
Reading
rate
Slow
Medium
V RMS
12.000 mV
100.000 mV
0.12000 V
1.00000 V
1.2000 V
10.0000 V
12.000 V
100.000 V
120.00 V
750.00 V
40.00 mV
360.00 mV
0.4000 V
3.6000 V
4.000 V
36.000 V
40.00 V
360.00 V
120.0 V
750.0 V
Table 4-
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
Input
frequency
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
Table 4-
8. (Be certain to allow for
U3402A User’s and Service Guide
Range
120 mV
120 mV
1.2 V
1.2 V
12 V
12 V
120 V
120 V
750 V
750 V
400 mV
400 mV
4 V
4 V
40 V
40 V
400 V
400 V
750 V
750 V
8. Provide the indicated
Error from nominal
one year
±0.124 mV
±0.3 mV
±1.24 mV
±3 mV
±12.4 mV
±30 mV
±124 mV
±0.3 V
±1.24 V
±2.5 V
±0.48 mV
±1.12 mV
±4.8 mV
±11.2 mV
±48 mV
±112 mV
±480 mV
±1.12 V
±4.2 V
±5.5 V