FLUKE-190-104/AM/S Fluke Electronics, FLUKE-190-104/AM/S Datasheet - Page 29

no-image

FLUKE-190-104/AM/S

Manufacturer Part Number
FLUKE-190-104/AM/S
Description
SCOPEMETER 4CH 100MHZ COLOR SCC
Manufacturer
Fluke Electronics
Series
ScopeMeter® 190r
Type
Handheldr

Specifications of FLUKE-190-104/AM/S

Bandwidth
100MHz
Channels
4
Display Type
LCD - Color
Features
DMM, USB Interface, Save, Recall, Record, Trend
Probe Type
10:1 Voltage Probe (1 Blue, 1 Green, 1 Red, 1 Gray)
Sampling Rate (per Second)
1.25GS/s
Input Impedance
1M
Rise Time (typ)
3.5nS
Voltage - Input (max)
CAT III 1000V, CAT IV 600V
Voltage - Supply
7.2V Battery, 17.8VDC Adapter
Includes
Alligator Clips, Battery Charger/Adapter, Case, Hook Clips, Probe Set, Software, Test Leads, USB Cable
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
Other names
614-1177
Using Average, Persistence and Glitch
Capture
Using Average for Smoothing Waveforms
To smooth the waveform, do the following:
1
2
3
4
F4
SCOPE
ENTER
Display the
Open the
menu.
Jump to Average:
Select On... to open the
menu.
WAVEFORM OPTIONS
SCOPE
key labels.
AVERAGE
You can use the average functions to suppress random or
uncorrelated noise in the waveform without loss of
bandwidth. Waveform samples with and without smoothing
are shown in Figure 9.
Smart average
In the normal average mode occasional deviations in a
waveform just distort the averaged wave shape, and do
not show up on screen clearly. When a signal really
changes, for instance when you probe around, it takes
quite some time before the new wave shape is stable.
With smart averaging you can quickly probe around, and
incidental waveform changes like a line flyback in video
show up on screen instantly.
5
6
Using Average, Persistence and Glitch Capture
ENTER
ENTER
Select Average factor: Average
64. This averages the outcomes
of 64 acquisitions.
Select Average: Normal (normal
average) or Smart (smart
average, see below)
21