FLUKE-190-104/AM/S Fluke Electronics, FLUKE-190-104/AM/S Datasheet - Page 31

no-image

FLUKE-190-104/AM/S

Manufacturer Part Number
FLUKE-190-104/AM/S
Description
SCOPEMETER 4CH 100MHZ COLOR SCC
Manufacturer
Fluke Electronics
Series
ScopeMeter® 190r
Type
Handheldr

Specifications of FLUKE-190-104/AM/S

Bandwidth
100MHz
Channels
4
Display Type
LCD - Color
Features
DMM, USB Interface, Save, Recall, Record, Trend
Probe Type
10:1 Voltage Probe (1 Blue, 1 Green, 1 Red, 1 Gray)
Sampling Rate (per Second)
1.25GS/s
Input Impedance
1M
Rise Time (typ)
3.5nS
Voltage - Input (max)
CAT III 1000V, CAT IV 600V
Voltage - Supply
7.2V Battery, 17.8VDC Adapter
Includes
Alligator Clips, Battery Charger/Adapter, Case, Hook Clips, Probe Set, Software, Test Leads, USB Cable
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
Other names
614-1177
Figure 10. Using Persistence to Observe Dynamic
Select Display: Dot-join: Off to
display measured samples only. Dot
join off may be useful when
measuring for example modulated
signals or video signals.
Select Display: Normal to turn the
envelope mode off and the dot-join
function on.
Signals
Displaying Glitches
To capture glitches on a waveform, do the following:
You can use this function to display events (glitches or
other asynchronous waveforms) of 8 ns (8 nanoseconds,
due to ADC’s with 125 MS/s sampling speed) or wider, or
you can display HF modulated waveforms.
When you select the 2 mV/div range Glitch Detect will
automatically be turned Off. In the 2 mV/div range you can
set Glitch Detect On manually.
1
2
3
4
Using Average, Persistence and Glitch Capture
F4
F4
SCOPE
ENTER
Display the
Open the
menu.
Select Glitch: On
Exit the menu.
WAVEFORM OPTIONS
SCOPE
key labels.
23