FLUKE-190-204/AM Fluke Electronics, FLUKE-190-204/AM Datasheet - Page 32

SCOPEMETER 4CH 200MHZ COLOR AMER

FLUKE-190-204/AM

Manufacturer Part Number
FLUKE-190-204/AM
Description
SCOPEMETER 4CH 200MHZ COLOR AMER
Manufacturer
Fluke Electronics
Series
ScopeMeter® 190r
Type
Handheldr

Specifications of FLUKE-190-204/AM

Bandwidth
200MHz
Channels
4
Display Type
LCD - Color
Features
DMM, USB Interface, Save, Recall, Record, Trend
Probe Type
10:1 Voltage Probe (1 Blue, 1 Green, 1 Red, 1 Gray)
Sampling Rate (per Second)
1.25GS/s
Input Impedance
1M
Rise Time (typ)
1.7nS
Voltage - Input (max)
CAT III 1000V, CAT IV 600V
Voltage - Supply
7.2V Battery, 17.8VDC Adapter
Includes
Alligator Clips, Battery Charger/Adapter, Case, Hook Clips, Probe Set, Software, Test Leads, USB Cable
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
Other names
614-1174
Fluke 190 Series II
Users Manual
Suppressing High Frequency Noise
Switching the glitch detection off (Glitch: Off) will
suppress the high frequency noise on a waveform.
Averaging will suppress the noise even more.
See also
on page 21.
Glitch capture and average do not affect bandwidth.
Further noise suppression is possible with bandwidth
limiting filters. See
page 27.
24
1
2
3
4
F4
Using Average for Smoothing Waveforms
SCOPE
ENTER
ENTER
Working with Noisy Waveforms
Display the
Open the
menu.
Select Glitch: Off, then select
Average: On… to open the
AVERAGE
Select Average 8 .
menu
WAVEFORM OPTIONS
SCOPE
key labels.
on
Acquiring Waveforms
Setting the Acquisition Speed and Waveform
Memory Depth
To set the acquisition speed, do the following:
1
2
3
F4
SCOPE
ENTER
Display the
Open the
menu.
Select Acquisition:
Fast – for fast trace update rate;
shortest record length, decreased
zoom rate, no readings possible.
Full – maximum waveform detail;
10,000 samples per trace record
length, maximum zoom rate,
lower trace update rate.
Normal – optimal trace update
rate and zoom range combination
WAVEFORM OPTIONS
SCOPE
key labels.