Description
4x1 unterminated multiplexer/50 characteristic impedance
Manufacturer
NI [National Instruments Corporation]
Datasheet
492
1.3 GHz and 4 GHz Multiplexers
Overview
The National Instruments PXI-2590 and PXI-2591 are 4x1 high-
frequency unterminated multiplexing switches. The NI SCXI-1190 and
NI SCXI-1191 perform the same functions as the PXI-2590 and
PXI-2591 but also offer a quad 4x1 multiplexing configuration. The
switches are capable of switching signals from DC to 1.3 GHz
(SCXI-1190 and PXI-2590) and DC to 4 GHz (SCXI-1191 and
PXI-2591). The characteristic impedance of the channels is 50 .
of the switches is 24 V
SCXI-1191 and the PXI-2591, the maximum voltage rating is 30 V
with a maximum current of 330 mA. All are well suited for
applications that require the routing of high-frequency signals inside
automated test equipment (ATE) systems because they can switch
signals with a very low insertion loss. In addition, the excellent
voltage standing-wave ratio (VSWR) and isolation parameters make
these modules the perfect choice in a system geared to high-
frequency applications as shown in Table 1.
Extended Features and Specifications
National Instruments switch modules are built with a number of core
features that are covered in detail in the Switch Overview section.
For additional information about the PXI-2590, PXI-2591, SCXI-1190, and
SCXI-1191, including software, certifications and compliance, relay
control, etc., please see page 20. For detailed specifications, please
see page 504.
National Instruments • Tel: (800) 433-3488 • Fax: (512) 683-9300 • info@ni.com • ni.com
For the SCXI-1190 and the PXI-2590, the maximum voltage rating
NI SCXI-1190, NI PXI-2590, NI SCXI-1191, NI PXI-2591
• 4x1 unterminated multiplexer
• 50
• High-bandwidth
• Excellent insertion loss, VSWR,
• Fully software programmable
SCXI-1190 (quad 4x1),
PXI-2590 (single 4x1)
• 1.3 GHz bandwidth
• SMB direct connectivity
• 1 A at 24 VDC capacity
SCXI-1191 (quad 4x1),
PXI-2591 (single 4x1)
• 4 GHz bandwidth
• SMA direct connectivity
• 330 mA at 30 VDC capacity
electromechanical relays
and isolation specifications
characteristic impedance
rms
with a maximum current of 1 A. For the
Operating Systems
• Windows 2000/NT/XP
Recommended Software
• LabVIEW
• LabVIEW Real-Time Module
• LabWindows/CVI
• Measurement Studio
• NI Switch Executive
Other Compatible Software
• Visual Basic
• C/C++
Driver Software (included)
• NI-SWITCH
Compliance
• CE
(PXI Modules only)
rms
Table 1. Parameters of High-Frequency Switching Modules
Module
SCXI-1190, PXI-2590 (@ 1.3 GHz)
SCXI-1190, PXI-2590 (@ 2 GHz)
SCXI-1191, PXI-2591 (@ 4 GHz)
Ordering Information
NI SCXI-1190 ................................................................776572-90
NI PXI-2590 ..................................................................777987-01
NI SCXI-1191 ................................................................776572-91
NI PXI-2591 ..................................................................778339-01
Includes switch module and NI-SWITCH driver software.
For information on extended warranty and value added
services, see page 20.
See page 499 for accessory and cable information.
BUY ONLINE!
Visit
and/or scxi1191.
ni.com/products
and enter pxi2590, pxi2591, scxi1190,
Insertion Loss
< 1.5 dB
< 3 dB
< 0.9 dB
VSWR
< 1.5
< 2.5
< 1.5

Related parts for NIPXI-2590

NIPXI-2590 Summary of contents

Page 1

GHz and 4 GHz Multiplexers NI SCXI-1190, NI PXI-2590, NI SCXI-1191, NI PXI-2591 • 4x1 unterminated multiplexer • 50 characteristic impedance • High-bandwidth electromechanical relays • Excellent insertion loss, VSWR, and isolation specifications • Fully software programmable SCXI-1190 (quad ...

Page 2

Switch Specifications Specifications (continued) 2 1.9 1.8 1.7 1.6 1.5 1.4 1.3 1.2 1 100 200 300 400 500 600 Frequency (MHz) Input Characteristics Maximum switching voltage............................ 150 V, CAT I Maximum switching current ............................ 0.5 A (per ...

Page 3

Specifications (continued) SCXI-1191, PXI-2591 Input Characteristics Number of channels per bank ......................... 4 Number of banks.............................................. 4 SCXI-1191 .................................................. 4 PXI-2591..................................................... 1 Input voltage Channel-to-ground..................................... 30 V /VDC rms Maximum switching voltage AC ............................................................ 30 V rms DC ............................................................ 30 ...

Related keywords