CRD5490-Z Cirrus Logic, CRD5490-Z Datasheet - Page 15

no-image

CRD5490-Z

Manufacturer Part Number
CRD5490-Z
Description
Power Management IC Development Tools CS5463 Pwr Meas/Mntr Reference Design
Manufacturer
Cirrus Logic
Type
Energy Meterr
Datasheet

Specifications of CRD5490-Z

Rohs
yes
Product
Reference Design Boards
Tool Is For Evaluation Of
CS5490
Input Voltage
90 VAC to 260 VAC
Output Current
15 A
CRD5490-Z
3.3.5
Calibration Tab
The Calibration tab is used to write and display to the CS5490 offset and gain calibration registers. The built-in
calibration sequences of the CS5490 that are used to set the calibration register can be initiated from the Calibration
tab. AC offset and gain calibration can be performed on the voltage channel, the current channel, or both
simultaneously. Refer to the CS5490 data sheet, entitled Two Channel Energy Measurement IC, and Application
Note AN366, entitled CS5480/84/90 Energy Measurement IC Calibration, for more details on calibration. The
Calibration tab is illustrated in Figure 14.
Figure 14. Calibration Tab
To access the Calibration tab, conversion must be stopped. Otherwise, the Quick Measurements tab will be
displayed with the error message shown in Figure 15.
Figure 15. Quick Measurements Tab with Error Message Displayed (Partially Illustrated)
The Refresh button updates the contents of the screen by reading all the register values from the CS5490. It is
recommended to click the Refresh button when entering the Calibration tab or after modifying any registers to reflect
the current status of the CS5490.