LTC1966IMS8 Linear Technology, LTC1966IMS8 Datasheet - Page 26

IC PREC RMS/DC CONV MCRPWR 8MSOP

LTC1966IMS8

Manufacturer Part Number
LTC1966IMS8
Description
IC PREC RMS/DC CONV MCRPWR 8MSOP
Manufacturer
Linear Technology
Datasheet

Specifications of LTC1966IMS8

Current - Supply
155µA
Voltage - Supply
2.7 V ~ 5.5 V
Mounting Type
Surface Mount
Package / Case
8-MSOP, Micro8™, 8-uMAX, 8-uSOP,
Lead Free Status / RoHS Status
Contains lead / RoHS non-compliant

Available stocks

Company
Part Number
Manufacturer
Quantity
Price
Part Number:
LTC1966IMS8
Manufacturer:
LT
Quantity:
10 000
Part Number:
LTC1966IMS8
Manufacturer:
LINEAR/凌特
Quantity:
20 000
Part Number:
LTC1966IMS8#TRPBF
Manufacturer:
LT
Quantity:
670
APPLICATIO S I FOR ATIO
LTC1966
The best frequency for the calibration signal is roughly ten
times the – 0.1% DC error frequency. For 1 F, – 0.1% DC
error occurs at 8Hz, so 80Hz is a good calibration frequency,
although anywhere from 60Hz to 100Hz should suffice.
The trade-off here is that on the one hand, the DC error is
input frequency dependent, so a calibration signal fre-
quency high enough to make the DC error negligible
should be used. On the other hand, as low a frequency as
can be used is best to avoid attenuation of the calibrated
AC signal, either from parasitic RC loading or insufficient
op amp gain. For instance, with a 1kHz calibration signal,
a 1MHz op amp will typically only have 60dB of open-loop
gain, so it could attenuate the calibration signal a full 0.1%.
AC-Only, 2 Point
The next most significant error for AC-coupled applica-
tions will be the effect of output offset voltage, noticeable
at the bottom end of the input scale. This too can be
calibrated out if two measurements are made, one with a
full-scale sine wave input and a second with a sine wave
input (of the same frequency) at 10% of full scale. The
trade-off in selecting this second level is that it should be
small enough that the gain error effect becomes small
compared to the gain error effect at full scale, while on the
other hand, not using so small an input that the input offset
voltage becomes an issue.
The calculations of the error terms for a 200mV full-scale
case are:
DC, 2 Point
DC-based calibration is preferable in many cases because
a DC voltage of known, good accuracy is easier to generate
than such an AC calibration voltage. The only down side is
26
Gain =
Output Offset =
Reading at 200mV – Reading at 20mV
U
Reading at 20mV
U
180mV
Gain
W
– 20mV
U
that the LTC1966 input offset voltage plays a role. It is
therefore suggested that a DC-based calibration scheme
check at least two points: full scale. Applying the –full-
scale input can be done by physically inverting the voltage
or by applying the same +full-scale input to the opposite
LTC1966 input.
For an otherwise AC-coupled application, only the gain
term may be worth correcting for, but for DC-coupled
applications, the input offset voltage can also be calcu-
lated and corrected for.
The calculations of the error terms for a 200mV full-scale
case are:
Note: Calculation of and correction for input offset voltage
are the only way in which the two LTC1966 inputs (IN1,
IN2) are distinguishable from each other. The calculation
above assumes the standard definition of offset; that a
positive offset is the case of a positive voltage error inside
the device that must be corrected by applying a like
negative voltage outside. The offset is referred to which-
ever pin is driven positive for the +full-scale reading.
DC, 3 Point
One more point is needed with a DC calibration scheme to
determine output offset voltage: +10% of full scale.
The calculation of the input offset is the same as for the
2-point calibration above, while the gain and output offset
are calculated for a 200mV full-scale case as:
Gain =
Input Offset =
Gain =
Output Offset =
Reading at 200mV +Reading at – 200mV – 400mV • Gain
Reading at 200mV + Reading at – 200mV
Reading at 200mV – Reading at 20mV
Reading at – 200mV – Reading at 200mV
180mV
400mV
2
2 •Gain
sn1966 1966fas

Related parts for LTC1966IMS8