MAX110EVC32-DIP Maxim Integrated Products, MAX110EVC32-DIP Datasheet - Page 5

no-image

MAX110EVC32-DIP

Manufacturer Part Number
MAX110EVC32-DIP
Description
EVAL KIT FOR MAX110
Manufacturer
Maxim Integrated Products
Datasheets

Specifications of MAX110EVC32-DIP

Number Of Adc's
1
Number Of Bits
14
Sampling Rate (per Second)
50
Data Interface
Serial
Inputs Per Adc
2 Differential
Input Range
±VREF
Power (typ) @ Conditions
4.35mW @ 50SPS
Voltage Supply Source
Single Supply
Operating Temperature
0°C ~ 70°C
Utilized Ic / Part
MAX110, MAX111
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
1) To monitor the MAX110 supply current, install a cur-
2) Follow the quick-start instructions to start the
3) From the Control Panel screen, press “S” to enter
4) From the Shutdown Power Cycling screen, press
5) From the Shutdown Power Cycling screen, press
A typical application may sample the MAX110 at regu-
lar intervals, keeping the MAX110 shut down between
samples. This type of usage can be evaluated by using
the “P” command in the Shutdown Power Cycling
screen. The power-cycling loop repeatedly puts the
MAX110 into sleep mode for a user-specified length of
time, wakes up the MAX110, takes a reading, and
powers the MAX110 down again. The “P” command
runs continuously, until halted by either the “U” or “D”
command.
The parameters that pertain to the “P” command are as
follows: the state of the PD and PDX bits before and
after shutdown, the sleep time, and the optional NOOP,
AUTO NULL, and GAIN CAL cycles.
The sampling rate is determined by setting the sleep
time, which is the length of time the program keeps the
MAX110 in shutdown mode. The program allows a
maximum sleep time of 65 seconds.
If the internal RC oscillator is used, activating the PDX
bit may cause the BUSY line to remain low. This condi-
tion should be cleared by starting the power-up
sequence with a configuration word whose MSB is 0 (a
NOOP cycle). If an external oscillator is used, the PDX
bit has no effect and the NOOP cycle is not required.
Supply current will not be as low as in the internal RC
oscillator mode, unless the external oscillator is halted
during shutdown.
The PDX power-down bit has no effect unless the
internal RC oscillator is being used. Normally, the PD
and PDX bits should be set to 1 in shutdown mode, and
both should be 0 in active mode.
rent meter before applying power to the MAX110 EV
kit. Refer to the
Typical supply current is less than 1mA during nor-
mal operation, and less than 4µA in shutdown mode
if the internal RC oscillator is used.
MAX110 software on the IBM PC.
the Shutdown Power Cycling screen.
“D” to shut down the MAX110. The supply current
should drop to less than 4µA.
“U” to power-up the MAX110. The supply current
should settle at less than 1mA. There will be short
current spikes whenever a command is sent to the
MAX110.
MAX110 Evaluation System/Evaluation Kit
_______________________________________________________________________________________
Measuring Supply Current section.
Jumper J9 can be used to measure the supply current
drawn by the MAX110 IC. To measure the supply current,
use the following procedure:
1) Exit the MAX110 software and then turn off the
2) Carefully cut the printed circuit board trace at
3) Connect a current meter between the two pins of
4) Turn on the MAX110 EV kit and restart the MAX110
5) Observe the supply current in both operating and
To restore the MAX110 EV kit, use the following
procedure:
1) Turn off the MAX110 EV kit power.
2) Disconnect the current meter.
3) Install a 2-pin header and jumper at J9, or use a
If the MAX110 V
current can be measured in a similar way. Remove the
shunt from J4 1-2 and connect a current meter. Current
will flow from J4 pin 2 to J4 pin 1. Or, the current meter
may be connected between the V
-5V edge pad. Current will flow from the V
to the -5V edge pad.
When starting the MAX110 program from the DOS
prompt, several command-line options are available.
For a list of available options, run MAX110 with the “?”
command-line option. Refer to Table 3.
When the MAX110 program begins operation, it is in its
Opening screen. Use the space bar to select the serial
communications port to which you have connected the
80C32 module. Press the ENTER key to advance to the
Terminal screen.
The MAX110 program displays its Terminal screen
when it is establishing communications with the 80C32
module. When power is applied to the 80C32 module
or reset is pressed, banner message identifying the
80C32 module is displayed. After the module com-
pletes its self check, it says that all tests have passed.
At this point, use the ALT+L command to load the RAM
resident program into the 80C32 module. After the RAM
resident program has been loaded, use the ALT+C
command to advance to the Control Panel screen.
__Detailed Description of Software
MAX110 EV kit power.
location J9.
J9. The direction of current flow is marked with an
arrow on the board.
program to evaluate the desired operating mode.
shutdown modes.
small piece of wire to reconnect J9.
SS
is connected to -5V, the -5V supply
Measuring Supply Current
SS
edge pad and the
SS
edge pad
5