85RF Fluke Electronics, 85RF Datasheet - Page 33

no-image

85RF

Manufacturer Part Number
85RF
Description
PROBE HIGH FREQ 500MHZ
Manufacturer
Fluke Electronics
Type
High Frequencyr
Datasheets

Specifications of 85RF

Impedance
10M Ohm
Voltage - Max
30VRMS, 200VDC
Lead Free Status / RoHS Status
Lead free by exemption / RoHS compliant by exemption
Ratings
-
Current Rating
-
Color
-
Termination
-
Wire Insulation Material
-
Cable Length
-
Other names
532457
2-11. AC/DC Voltage (V)
Selection of the ac or dc voltage (V) functions is described in Figure 2-6.
The 8062A offers five ac and five dc voltage ranges: 200 mV, 2V, 20V,
200V, and 750V ac/1000V dc. All ranges present a 10 M input impedance,
which is shunted by <100 pF.
2-12. True RMS Measurement
One of the most useful features of the 8062A is the direct measurement of
true rms or effective ac voltages and ac currents. Mathematically, rms is
defined as the square root of the sum of the squares of the ac and dc
components. In physical terms, rms is equivalent to the dc value that
dissipates the same amount of heat in a resistor as the original waveform.
The reason that rms is so valuable is that it greatly simplifies the analysis of
1. Select a range.
2. Set AC/DC switch out
3. Press switch in to select
4. Ensure all other switches are out.
5. Connect the test leads as shown above.
6. Heed the input overload limits (Table 2-2) and connect the leads
7. Read the measured value on the display.
Voltage (V)
for DC, in for AC.
voltage function.
to the circuit being measured.
Figure 2-6. Voltage Operation
Operation Instructions
Operation
Low (-)
High (+)
dy08f.eps
2-11