85RF Fluke Electronics, 85RF Datasheet - Page 41

no-image

85RF

Manufacturer Part Number
85RF
Description
PROBE HIGH FREQ 500MHZ
Manufacturer
Fluke Electronics
Type
High Frequencyr
Datasheets

Specifications of 85RF

Impedance
10M Ohm
Voltage - Max
30VRMS, 200VDC
Lead Free Status / RoHS Status
Lead free by exemption / RoHS compliant by exemption
Ratings
-
Current Rating
-
Color
-
Termination
-
Wire Insulation Material
-
Cable Length
-
Other names
532457
Resistance measurements for all ranges are made using a two-wire
ratiometric technique. This means that test lead resistance may affect the
accuracy in the 200 range. You can correct this error by shorting the test
leads together, reading the test lead resistance, and then subtracting it from
resistance readings. The most convenient way to do this is with the relative
function as described later in this chapter. This technique is also useful for
removing the .02 error factor in the 200 range (refer to resistance
specifications in Chapter 1).
2-18. Autoranging Megohms
When the autoranging M range is selected, the 8062A automatically selects
the range appropriate for the measurement. The measurement resolution
decreases in the two higher M ranges as shown in Table 2-3. Readings
made at the crossover points between ranges are microcomputer-stabilized
by an offset in the upscale and downscale directions. Range changes are
made at 2.00 M and 20.00 M as readings go upscale, or at 19.0 M and
1.90 M as readings go downscale.
Autorange
Autorange
M
k
Table 2-3. Resistance Function Autoranges and Resolution
2 M
20 M
300 M
2 k
20 k
300 k
Range
100 to 300 M
20 to 99.9 M
100 to 299 k
20 to 99.9 k
Resolution
Operation Instructions
100 k
10 k
100
1 M
100
0.1
1 k
10
No. of Digits
Operation
Possible in
Reading
3
3
3
3
2-19
2