80PK-25 Fluke Electronics, 80PK-25 Datasheet

no-image

80PK-25

Manufacturer Part Number
80PK-25
Description
PROBE PIERCING TEMP SUREGRIP
Manufacturer
Fluke Electronics
Series
SureGrip™r
Type
K-Type Thermocoupler
Datasheets

Specifications of 80PK-25

Usage
Liquids, Gels
Tip - Type
Immersion / Penetration, Pointed
Plug Type
Sub-Mini (Thermocouple Plug)
Probe Temperature Range
-40 ~ 662°F (-40 ~ 350°C)
Cable Length
39.370" (1000.00mm)
Cable Insulation
Poly-Vinyl Chloride (PVC)
Plug Color
Yellow
Probe Length
4.000" (101.60mm)
Probe Material
Stainless Steel
For Use With
614-1037 - THERMOMETER HVACPRO IR
Lead Free Status / RoHS Status
Lead free by exemption / RoHS compliant by exemption
Other names
2098706
Introduction
The 80PK-25 SureGripTM Piercing Temperature Probe is
designed for use in liquids, gels, soft, and semi-hard
materials. The probe is made of 316 stainless steel, which
is suitable for food service use. The thermocouple junction
of the probe is protected from tip to handle by the stainless
steel sheath. The 40-inch (1-meter) cable is terminated
with a Type K miniature thermocouple connector with
0.792-mm (.312-inch) pin spacing. The 80PK-25 can be
used with any temperature-measuring instrument that is
designed to accept Type K thermocouples and has a
miniature connector input.
Specifications
Type
K,
Measurement Range of Stainless Steel Probe
-40 °C to 350 °C (-40 °F to 662 °F)
Accuracy
(With respect to ANSI MC96.1 Standard Limits of Error)
November 2004 Rev.1, 8/07
© 2004-2007 Fluke Corporation. All rights reserved.
All product names are trademarks of their respective companies
-40 °C to 0 °C
(-40 °F to 32 °F)
0 °C to 260 °C
(32 °F to 500 °F)
Standard grade Ni-Cr vs Ni-Al (Chromel vs Alumel)
To avoid electrical shock, do not use this
probe when voltages exceeding 24 V ac rms
or 60 V dc are present. The probe tip is
electronically connected to the output
terminals.
All error calculations should be done in °C, then
scaled to °F.
Range
Temperature Probe
SureGrip
XWWarning
Note
±2.2 °C
±1.1 °C
Instruction Sheet
80PK-25
(% of reading)
TM
Accuracy
Piercing
®

80PK-25 Summary of contents

Page 1

... The 40-inch (1-meter) cable is terminated with a Type K miniature thermocouple connector with 0.792-mm (.312-inch) pin spacing. The 80PK-25 can be used with any temperature-measuring instrument that is designed to accept Type K thermocouples and has a miniature connector input. ...

Page 2

Output @ 25 °C (77 °F) =1.00 mV (reference junction at 0 °C) Seebeck Coefficient @ 25 °C (77 °F) = 40.50 μV / °C Measurement Time (Time Constant): 3.0 seconds in 100 °C still water at sea level pressure ...

Page 3

... This is the maximum temperature limitation of the Hytrel handle. See Food Service Limitations. Exposure Limitations The sheath material of the 80PK-25 is stainless steel. It should not be exposed to halides or sulfides. Food Service Limitations The 316 stainless steel probe is suitable for contact with hot or cold foods within the specified temperature range of the probe ...

Page 4

... The ohmmeter should read less if there is continuity. Scale Conversions Use the following equation to convert °C to °F: Use the following equation to convert °F to °C: (° °F (°F –32) x 0.5556 = °C Figure 1. 80PK-25 bae01f.eps ...

Page 5

Contacting Fluke For application or operation assistance, or information on Fluke products, call: USA: 1-888-44-FLUKE (1-888-443-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Europe: +31 402-675-200 Japan: +81-3-3434-0181 Singapore: +65-738-5655 Anywhere in the world: +1-425-446-5500 For USA Service: 1-888-99-FLUKE (1-888-993-5853) Or, visit Fluke's Web ...