TMD-10 Amprobe, TMD-10 Datasheet - Page 8

THERMOMETER DUAL CHANNEL

TMD-10

Manufacturer Part Number
TMD-10
Description
THERMOMETER DUAL CHANNEL
Manufacturer
Amprobe
Type
Handheldr
Datasheets

Specifications of TMD-10

Temperature Range
32 ~ 122°F (0 ~ 50°C)
Temperature Display
C°/F°
Display Type
LCD
Input Type
Thermocouples (2)
Features
Hold, MIN/MAX, T1/T2/T1-T2
Probe Type
J, K
Battery Type
9 Volt
Includes
Case, Thermocouple - K-Type Bead Probe (2)
Lead Free Status / RoHS Status
Not applicable / Not applicable
Other names
3052394
To use the Relative mode for T2, disable
> 1 second to swap T2 to the top display and T1 to bottom display.
T1 T2 / T1-T2
T1 T2 displays the 2 inputs independently. Press and hold
second to swap display location of the display location of T1 and T2.
(T1-T2)
Press
shows the T1-T2 temperature. The lower display shows the top display
selected in the T1 T2 mode. Press
Auto Power off (APo)
The TMD-10 powers down automatically after approximately 15 minutes of
inactivity to save battery life. Press the � button to turn the TMD-10 ON. You
can disable Auto Power Off by pressing the
TMD-10 is ON.
� The power cannot be turned oFF in the oFFSET SETUP mode.
Thermocouple Error Correction (oFS, offset)
User can add an offset to compensate for the error of specific thermocouple
probes.
1.
2.
3.
4.
5.
6.
7.
Press and hold the
mode.
Immerse the thermocouple(s) in thermos bath with crushed ice and fresh
water (0.0 °C / 32.0 °F).
Press the
Offset value. The allowable adjustment range is ± 5 °C / ± 9 °F in 0.1°C /°F
increments.
Press
will remain on the LCD.
Press and hold
T1 to bottom display. Press
Repeat steps 1, 2, 3 and 4 for T2.
To remove the Offset, Repeat steps 1, 2, 3 and 4 setting the offset (lower
display) to 0.0. When both thermocouple offset are zeroed out, the
“Offset” icon will be turned off.
button to display T1 minus T2 temperature value. The top display
(SET) button to save and leave setup mode. The “Offset” icon
button to increase or power � button to decrease the
button > 1 second to swap T2 to the top display and
(OFS) button > 1 second to enter the Offset Setup
button to switch T2 from type K to type J.
7
button again to disable this function
and press and hold
button > 1 second after the
button > 1
key