TMD-10 Amprobe, TMD-10 Datasheet - Page 9

THERMOMETER DUAL CHANNEL

TMD-10

Manufacturer Part Number
TMD-10
Description
THERMOMETER DUAL CHANNEL
Manufacturer
Amprobe
Type
Handheldr
Datasheets

Specifications of TMD-10

Temperature Range
32 ~ 122°F (0 ~ 50°C)
Temperature Display
C°/F°
Display Type
LCD
Input Type
Thermocouples (2)
Features
Hold, MIN/MAX, T1/T2/T1-T2
Probe Type
J, K
Battery Type
9 Volt
Includes
Case, Thermocouple - K-Type Bead Probe (2)
Lead Free Status / RoHS Status
Not applicable / Not applicable
Other names
3052394
SPECIFICATIoNS
general
Display: Dual 5-digit LCD displays
Overrange indication:
Low battery: the “�"sign will appear while battery is low.
Power Supply: 9V NEDA 1604, IEC 6F22, JIS 006P battery.
Battery life: approx. 200 hours.
Auto power off: approx. 15 minutes.
Environment: Indoor operation, - Altitude: Up to 2000 m.
Temperature / Humidity:
Dimensions: 130 x 56 x 38 mm (5.1 x 2.2 x 1.5 in.)
Weight: 170g( 0.37 lb.) battery included
following European Community Directives: 89/336/EEC (Electromagnetic
Compatibility) and 73/23/EEC (Low Voltage) as amended by 93/68/EEC (CE
Marking). However, electrical noise or intense electromagnetic fields in the
vicinity of the equipment may disturb the measurement circuit. Measuring
instruments will also respond to unwanted signals that may be present within
the measurement circuit. Users should exercise care and take appropriate
precautions to avoid misleading results when making measurements in the
presence of electronic interference.
Electrical (23 °C ± 5°C, < 80 % RH)
Temperature scale: °C or ° F
Thermocouple type:
K-type: -200°C to 1370 °C (-328 °F to 2498 °F)
J-type: -200°C to 1050 °C (-328°F to 1922 °F)
Resolution: 0.1
Accuracy: ± (0.05 % rdg + 0.7 °C), ± (0.05 % rdg + 1.4 °F)
Input protection: 24V DC or AC.
-EMC: EN61326-1 This product complies with requirements of the
“OL” stands for positive temperature
“-OL” stands for negative temperature.
Operating: 0°C to 50°C (32°F to 122°F); < 80% RH
Storage: 0°C to 50°C (32°F to 122°F); < 80% RH
8