U2001A AGILENT TECHNOLOGIES, U2001A Datasheet - Page 14

USB POWER SENSOR, 6GHZ, -60DBM To +20DBM

U2001A

Manufacturer Part Number
U2001A
Description
USB POWER SENSOR, 6GHZ, -60DBM To +20DBM
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of U2001A

Frequency Measuring Range
10MHz To 6GHz
Amplitude Accuracy
± 3%
Standing Wave Ratio
1.19
Connector Type
N
External Height
35.9mm
External Width
46mm
External Depth
163.75mm
Lead Free Status / RoHS Status
na
Specifications (continued)
Settling time
Manual filter, 10-dB decreasing power step (not across switching points)
Auto filter, default resolution, 10-dB decreasing power step
Typical
settling
times
Settling time with auto filter, default resolution, and a 10 dB decreasing power step (not across the switching point)
Settling time = 25 ms*
*
Number of averages
Settling time (s)
When a power step crosses through the sensor’s auto-range switching point, add 25 ms
Normal speed
x2 speed
450 ms
speed
45 ms
82 ms
1.3 s
1.5 s
1.6 s
20 s
24 s
24 s
x2
Normal
460 ms
speed
45 ms
90 ms
1
0.045
0.042
2.6 s
2.7 s
2.8 s
39 s
42 s
42 s
2
0.09
0.05
U2000/1/2/4A
+10 dBm
–10 dBm
–20 dBm
–30 dBm
–40 dBm
–50 dBm
+2 dBm
–4 dBm
4
0.17
0.09
8
0.34
0.17
U2000/1/2H
+20 dBm
+12 dBm
–10 dBm
–20 dBm
–30 dBm
–40 dBm
+6 dBm
0 dBm
14
16
0.66
0.34
32
1.3
0.66
U2000/1B
+40 dBm
+32 dBm
+26 dBm
+20 dBm
+10 dBm
–10 dBm
–20 dBm
0 dBm
64
2.6
1.3
128
5.2
2.6
Maximum dBm
Minimum dBm
High power path
Low power path
Sensor
dynamic
range
256
10.4
5.2
512
20.9
10.4
1024
41.9
20.9