FLUKE-FP PLUS Fluke Electronics, FLUKE-FP PLUS Datasheet - Page 14

no-image

FLUKE-FP PLUS

Manufacturer Part Number
FLUKE-FP PLUS
Description
THERM FOODPRO PLUS FOOD SAFETY
Manufacturer
Fluke Electronics
Series
FoodPro Plusr
Type
Handheld, Gunr
Datasheet

Specifications of FLUKE-FP PLUS

Temperature Range
-40 ~ 392°F (-40 ~ 200°C)
Temperature Display
C°/F°
Display Type
LCD
Input Type
Infrared; Probe
Features
Backlight, Hold, IP54 Sealed, LED Sight, MAX Reading; Timer
Probe Type
RTD
Battery Type
9 Volt
Includes
Pouch
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Other names
2477950
FoodPro Plus
Users Manual
contents of delicases, chilled salads and desserts, holding ovens,
rotisseries, etc.
Field Verification of Accuracy
Infrared (IRT) and Probe channels (RTD)
Follow the steps below to verify the accuracy of your food safety
thermometer. The 0 C (32 F) “stirred ice bath” is the recommended
verification reference point. Since it is more difficult to measure the surface
of hot water, the hot verification should be used only as a general accuracy
check of the IRT channel.
Cold Temperature Verification Check
10
If any questionable temperatures are indicated or if temperature
readings are within ±1 °C (±2 °F) of the HACCP “Danger Zone”
boundary, use the probe to check internal temperatures.
1.
2.
3.
Fill a large styrofoam cup halfway to the surface with ice cubes.
Add cold water to just below the rim of the cup.
Immerse the tip of a known calibrated probe thermometer
(reference probe) into the water and stir the mixture with the
probe for one minute, or until the probe temperature stabilizes.
Continue stirring the water with a straw or swizzle stick while
taking simultaneous temperature measurements with the
reference probe and the IR thermometer. The unit should be
held within 3 inches of the surface of the water (Figure 8). To
insure measurement accuracy the probe tip must be immersed
to a minimum depth of 12 mm (1/2 ").
Note