72-6851 TENMA, 72-6851 Datasheet - Page 21

POWER SUPPLY, BENCH, 35V

72-6851

Manufacturer Part Number
72-6851
Description
POWER SUPPLY, BENCH, 35V
Manufacturer
TENMA
Datasheet

Specifications of 72-6851

Power Supply Output Type
Variable
No. Of Outputs
1
Output Voltage
35V
Output Current
10A
Load Regulation
0.01%
Length
350.52mm
Width
210.82mm
Height
132.08mm
Output Type
Variable
Calibrated
No
Rohs Compliant
NA
Lead Free Status / RoHS Status
na
RS232 Remote Command Formats
GPIB Remote Command Formats
Tenma 72-6851 & 72-6853 Instruction manual
20
Serial input to the instrument is buffered in a 256 byte input queue which is filled, under interrupt,
in a manner transparent to all other instrument operations. The instrument will send XOFF when
approximately 200 characters are in the queue. XON will be sent when approximately 100 free
spaces become available in the queue after XOFF was sent. This queue contains raw (un-
parsed) data which is taken, by the parser, as required. Commands (and queries) are executed in
order and the parser will not start a new command until any previous command or query is
complete. In non–addressable RS232 mode responses to commands or queries are sent
immediately; there is no output queue. In addressable mode the response formatter will wait
indefinitely if necessary, until the instrument is addressed to talk and the complete response
message has been sent, before the parser is allowed to start the next command in the input
queue.
Commands must be sent as specified in the commands list and must be terminated with the
command terminator code 0AH (Line Feed, LF). Commands may be sent in groups with
individual commands separated from each other by the code 3BH (;). The group must be
terminated with command terminator 0AH (Line Feed, LF).
Responses from the instrument to the controller are sent as specified in the commands list. Each
response is terminated by 0DH (Carriage Return, CR) followed by 0AH (Line Feed, LF).
<WHITE SPACE>
which are specified as Addressable RS232 (ARC) control codes.
<WHITE SPACE>
The high bit of all characters is ignored.
The commands are case insensitive.
GPIB input to the instrument is buffered in a 256 byte input queue which is filled, under interrupt,
in a manner transparent to all other instrument operations. The queue contains raw (un-parsed)
data which is taken, by the parser, as required. Commands (and queries) are executed in order
and the parser will not start a new command until any previous command or query is complete.
There is no output queue which means that the response formatter will wait, indefinitely if
necessary, until the instrument is addressed to talk and the complete response message has
been sent, before the parser is allowed to start the next command in the input queue.
Commands are sent as
or more
elements.
A
A
<PROGRAM MESSAGES>
be any of the following:
Responses from the instrument to the controller are sent as
<RESPONSE MESSAGE>
MESSAGE TERMINATOR>
A
<PROGRAM MESSAGE UNIT>
<PROGRAM MESSAGE UNIT SEPARATOR>
<RESPONSE MESSAGE TERMINATOR>
NL
NL^END The new line character with the END message
^END
<PROGRAM MESSAGE UNIT>
The new line character (0AH)
is defined as character codes 00H to 20H inclusive with the exception of those
is ignored except in command identifiers. e.g. '*C LS' is not equivalent to '*CLS'.
The END message with the last character of the message
consists of one
.
are separated by
<PROGRAM MESSAGES>
is any of the commands in the remote commands list.
elements separated by
www.tenma.com
is the new line character with the END message NL^END.
<RESPONSE MESSAGE UNIT>
<PROGRAM MESSAGE TERMINATOR>
is the semi-colon character ';' (3BH).
by the controller, each message consisting of zero
Remote Commands
<PROGRAM MESSAGE UNIT SEPARATOR>
<RESPONSE MESSAGES>
followed by a
elements which may
<RESPONSE
. A