72-6851 TENMA, 72-6851 Datasheet - Page 4

POWER SUPPLY, BENCH, 35V

72-6851

Manufacturer Part Number
72-6851
Description
POWER SUPPLY, BENCH, 35V
Manufacturer
TENMA
Datasheet

Specifications of 72-6851

Power Supply Output Type
Variable
No. Of Outputs
1
Output Voltage
35V
Output Current
10A
Load Regulation
0.01%
Length
350.52mm
Width
210.82mm
Height
132.08mm
Output Type
Variable
Calibrated
No
Rohs Compliant
NA
Lead Free Status / RoHS Status
na
Tenma 72-6851 & 72-6853 Instruction manual
DIGITAL INTERFACES
OUTPUT SPECIFICATIONS - REMOTE OPERATION
RESPONSE TIME OVER RS232/GPIB:
GENERAL
Note: All voltage and current levels set via the keyboard are displayed on a separate 0.3” 4-digit
display. This entry preview system ensures that the user can observe the value entered before it
is effected thus avoiding possible error. The display is also used for setting additional functions.
When the output switch is on and no other function is selected, the display shows output power in
Watts.
Operational Functions:
RS232:
GPIB:
Output Voltage Setting:
Output Current Setting:
Setting Accuracy:
Output Switch:
Readback Resolution:
Readback Accuracy:
Interface:
Power Supply:
AC Input Voltage:
Power Consumption:
Operating Range:
Storage Range:
Environmental:
Size:
Weight:
Safety:
EMC:
Set voltage, set current, set OVP; set output on/off; read output
voltage; read output current; read output power.
Variable baud rate, 9600 baud maximum, 9 pin D-connector (female).
Can be controlled as a single instrument or as part of an Addressable
RS232 Chain (ARC).
Conforming with IEEE488.1 and IEEE488.2.
12 bit resolution (10mV steps)
12 bit resolution (10mA steps)
Voltage: ± (0.1% + 10mV); Current: ± (0.2% + 20mA)
Electronic by interface command.
Voltage: 10mV over the entire range.
Current: 10mA over the entire range.
Voltage: ± (0.2% of reading + 1 digit)
Current: ± (0.5% of reading + 1 digit)
<15ms (single command, buffer empty).
An internal time constant, T, (typically 22ms) governs the settling time
of a step voltage increase. Settling time to within 1% of the step
change = 4.6T, to 0.1% = 6.9T, to 0.01% = 9.2T; for example, after a
10V step the output will be within 1 digit (10mV = 0.1%) of its new
value in typically 150ms. For load current of 1 Amp or more, settling
times for downward steps will be very similar; however, response
times will be longer at low loads.
115V AC ±14%, 50/60Hz. Installation Category II.
700VA max.
5°C to +40°C, 20% to 80% RH.
−40°C to +70°C.
Indoor use at altitudes up to 2000m, Pollution Degree 2.
200 x 140 x 350mm (W x H x D); half rack width x 3U height;
(optional rack mounting kit available).
5.5kg.
Complies with IEC 1010-1
Complies with EN50081-1 and EN50082-1