72-6851 TENMA, 72-6851 Datasheet - Page 25

POWER SUPPLY, BENCH, 35V

72-6851

Manufacturer Part Number
72-6851
Description
POWER SUPPLY, BENCH, 35V
Manufacturer
TENMA
Datasheet

Specifications of 72-6851

Power Supply Output Type
Variable
No. Of Outputs
1
Output Voltage
35V
Output Current
10A
Load Regulation
0.01%
Length
350.52mm
Width
210.82mm
Height
132.08mm
Output Type
Variable
Calibrated
No
Rohs Compliant
NA
Lead Free Status / RoHS Status
na
Tenma 72-6851 & 72-6853 Instruction manual
∗SRE <nrf>
∗SRE?
∗STB?
∗SAV <nrf>
∗TST?
∗WAI
24
Sequential command.
Operation complete message generated immediately after execution.
Set the Service Request Enable Register to <nrf>. If the value of <nrf>, after rounding,
is less than 0 or greater than 255 an execution error is generated and error number 119
(Value out of range) is placed in the Execution Error Register.
Sequential command.
Operation complete message generated immediately after <rmt> is sent.
Returns the value of the Service Request Enable Register in <nr1> numeric format. The
syntax of the response is
Example. If the Service Request Enable Register contains 01000001b the
Sequential command.
Operation complete message generated immediately after <rmt> is sent.
Returns the value of the Status Byte Register in <nr1> numeric format. The syntax of
the response is
Example. If the Status Byte Register contains 01000001b the response to ∗STB?
Sequential command.
Operation complete message generated immediately after execution.
Save settings to store number <nrf>. The instrument settings are saved in the store
pointed to by <nrf>. If the value of <nrf>, after rounding, is less than 1 or greater than 25
an execution error is generated and error number 115 (illegal store number) is placed in
the Execution Error Register.
The scope of the ∗SAV command is as follows:
The output is set to stored values of Volts, current, OVP, delta V and delta A.
The output status is set to the stored value.
Sequential command.
Operation complete message generated immediately after <rmt> is sent.
Examine the state of the FLT bit in the Status byte register and return the result. If the
bit is clear the response is
any failure is indicated by a non-zero FLT bit and the response will be
The value in the FLT bit is 1 if an execution error 002 has been reported since the
instrument was powered up.
Sequential command.
Operation complete message generated immediately after execution.
Wait for operation complete true. As all commands are completely executed before the
next is started this command takes no additional action.
response to ∗SRE? will be 65<rmt>.
will be 65<rmt>.
www.tenma.com
<nr1><rmt>
<nr1><rmt>
0<rmt>
1<rmt>