DTX-MFM2 FLUKE NETWORKS, DTX-MFM2 Datasheet - Page 11

DTX-MFM2 REPLACEMENT FOR DTX-MFM

DTX-MFM2

Manufacturer Part Number
DTX-MFM2
Description
DTX-MFM2 REPLACEMENT FOR DTX-MFM
Manufacturer
FLUKE NETWORKS
Datasheet

Specifications of DTX-MFM2

Multimode Fiber Module Set
ROHS COMPLIANT
When using the fiber modules, use proper
cleaning procedures to clean all fiber connectors
before every use. Neglecting this step or using
improper procedures can cause unreliable test
results and may permanently damage the
connectors. See page 10.
Use only high-quality test reference cords that
comply with ISO/IEC 14763-3. Test the cords at
regular intervals. See Appendix A.
Cover all connectors with dust caps when not in
use.
Store the connector adapters for the fiber module
in the canisters provided.
Do not touch the photodiode lens (see Figure 4).
Do not overtighten the adapter or use tools to
tighten the adapter.
Use a Fluke Networks FiberInspector Video
Microscope to periodically inspect the fiber
module’s OUTPUT connector for scratches and
other damage.
Safety Information
5