DS1922T-F5# Maxim Integrated Products, DS1922T-F5# Datasheet - Page 23

IBUTTON TEMP LOGGER

DS1922T-F5#

Manufacturer Part Number
DS1922T-F5#
Description
IBUTTON TEMP LOGGER
Manufacturer
Maxim Integrated Products
Series
iButton®r
Datasheet

Specifications of DS1922T-F5#

Rohs Information
IButton RoHS Compliance Plan
Memory Size
512B
Memory Type
NVSRAM (Non-Volatile SRAM)
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
The fastest way to determine whether a programmed
temperature threshold was exceeded during a mission
is by reading the Alarm Status register. In a networked
environment that contains multiple DS1922L/
DS1922T iButtons, the devices that encountered an
alarm can quickly be identified by means of the
Conditional Search command (see the 1-Wire ROM
Function Commands section). The temperature alarm
only occurs if enabled (see the Temperature Sensor
Alarm section). The BOR alarm is always enabled.
Bit 7: Battery-On Reset Alarm (BOR). If this bit reads
1, the device has performed a power-on reset. This
indicates that the device has experienced a shock big
enough to interrupt the internal battery power supply.
The device can still appear functional, but it has lost its
factory calibration. Any data found in the data-log mem-
ory should be disregarded.
Bit 1: Temperature High Alarm Flag (THF). If this bit
reads 1, there was at least one temperature conversion
during a mission revealing a temperature equal to or
higher than the value in the Temperature High Alarm
register. A forced conversion can affect the THF bit.
This bit can also be set with the initial alarm in the SUTA
= 1 mode.
Bit 0: Temperature Low Alarm Flag (TLF). If this bit
reads 1, there was at least one temperature conversion
during a mission revealing a temperature equal to or
lower than the value in the Temperature Low Alarm reg-
ister. A forced conversion can affect the TLF bit. This bit
can also be set with the initial alarm in the SUTA = 1
mode.
Note: There is only read access to this register. Bits 4 to 6 have no function. They always read 1. Bits 2 and 3 have no function with
the DS1922L/DS1922T. They always read 0. The alarm status bits are cleared simultaneously when the Clear Memory Function is
invoked. See the Memory and Control Function Commands section for details.
Note: There is only read access to this register. Bits 0, 2, 5, 6, and 7 have no function.
ADDRESS
ADDRESS
0214h
0215h
BIT 7
BIT 7
BOR
______________________________________________________________________________________
1
Temperature Logger iButton with 8KB
BIT 6
BIT 6
1
1
Alarm Status
BIT 5
BIT 5
1
0
WFTA
BIT 4
BIT 4
1
The information in the General Status register tells the
host computer whether a mission-related command
was executed successfully. Individual status bits indi-
cate whether the DS1922L/DS1922T are performing a
mission, waiting for a temperature alarm to trigger the
logging of data, or whether the data from the latest mis-
sion has been cleared.
Bit 4: Waiting for Temperature Alarm (WFTA). If this
bit reads 1, the Mission Start Upon Temperature Alarm
was selected and the Start Mission command was suc-
cessfully executed, but the device has not yet experi-
enced the temperature alarm. This bit is cleared after a
temperature alarm event, but is not affected by the
Clear Memory command. Once set, WFTA remains set
if a mission is stopped before a temperature alarm
occurs. To clear WFTA manually before starting a new
mission, set the high temperature alarm (address
0209h) to -40°C and perform a forced conversion.
Bit 3: Memory Cleared (MEMCLR). If this bit reads 1,
the Mission Timestamp, Mission Samples Counter, and
all the alarm flags of the Alarm Status register have
been cleared in preparation of a new mission.
Executing the Clear Memory command clears these
memory sections. The MEMCLR bit returns to 0 as soon
as a new mission is started by using the Start Mission
command. The memory must be cleared for a mission
to start.
Bit 1: Mission in Progress (MIP). If this bit reads 1, the
device has been set up for a mission and this mission is
still in progress. The MIP bit returns from logic 1 to logic
0 when a mission is ended. See the Start Mission with
Password and Stop Mission with Password sections.
MEMCLR
BIT 3
BIT 3
0
Data-Log Memory
General Status Register Bitmap
Alarm Status Register Bitmap
BIT 2
BIT 2
0
0
BIT 1
BIT 1
THF
MIP
General Status
BIT 0
BIT 0
TLF
0