Service Information
Probe Performance
12
WARNING. To avoid personal injury and damage to the probe, remove the probe
from any signal source before attempting to adjust or service the probe.
This probe should be serviced only by qualified service personnel.
Do not service electrical equipment alone. If you must service or adjust
equipment that is attached to a signal or power source, do so only when another
person capable of rendering first aid and resuscitation is present.
This section provides procedures to check the performance of the P6021 or to
calibrate it. These procedures require the equipment listed in Table 6 on page 13.
Specifications given are the minimum necessary for accuracy. If equipment is
substituted, it must meet or exceed the specifications of the recommended
equipment. Test equipment is assumed to be correctly calibrated and operating
within the given specifications.
Also, if equipment is substituted, control settings or equipment setup may need
to be altered. For detailed operating instructions for the test equipment, refer to
the instruction manual for each unit.
To ensure measurement accuracy, check the performance of the probe and
termination whenever you begin using them with a different oscilloscope input,
especially when you have changed input capacitance. Recalibrate the probe if
necessary.
The recommended calibration interval of the probe and termination is every
twelve months.
Before calibrating the probe, thoroughly inspect and clean it as described in the
section entitled Maintenance on page 23. Dirty or worn mating surfaces between
the transformer and the lid can degrade low-frequency response. Clean these
surfaces if necessary.
Calibration procedures can also be used as performance checks by completing all
steps except those that adjust the probe. This checks the probe and termination to
the original performance standards without requiring you to remove the
termination cover or make internal adjustments.
P6021 Instruction Manual