Service Information
Performance Verification
14
Check Midband Accuracy. Refer to Figure 5 when making equipment connections.
Figure 5: Midband accuracy test setup
1. Set the P6021 termination sensitivity to 2 mA/mV.
2. Connect the Leveled Sinewave Output of the calibration generator to a
3. Set the DMM to ACV (autorange).
4. Set the Leveled Sinewave Output of the calibration generator to 50 kHz and
5. Measure and record the DMM output as M1.
6. Disconnect the coax cable from the BNC-to-dual banana adapter.
banana adapter
Termination for
BNC-to-dual
P6021 probe
BNC “T” connector. Connect one branch of the BNC “T” connector to the
calibration fixture. Connect the other branch of the BNC “T” to the DMM
input using a BNC-to-dual banana adapter.
an amplitude of 5 V
Step 7
DMM
P6021
probe
+ --
pp
.
Step 2, Step 6
connection
Coax cable
Calibration
BNC “T”
fixture
Calibration
generator
Sinewave
Leveled
Output
P6021 Instruction Manual
Note: If using a Wavetek
9100, use the Signal Out
BNC connector on the
rear of the instrument.