Check High Frequency Bandwidth. Refer to Figure 6 when making equipment
connections.
Figure 6: High frequency bandwidth test setup
1. Set the P6021 termination sensitivity to 2 mA/mV.
2. Connect the calibration fixture to the Leveled Sinewave Output of the
3. Connect the P6021 termination and probe to CH1 of the oscilloscope.
4. Set the oscilloscope Vertical Deflection to 10 mV/div.
5. Set the oscilloscope Horizontal Scale to 10 s/div.
6. Set the oscilloscope Auto Measurement to display “Amplitude”.
7. Set the Leveled Sinewave Output of the calibration generator to 50 kHz and
8. Connect the P6021 probe to the calibration fixture.
9. Enable the Leveled Sinewave Output of the calibration generator.
10. Measure and record the oscilloscope “amplitude” Auto Measurement as M1
11. Set the oscilloscope Horizontal Scale to 10 ns/div.
12. Set the Leveled Sinewave Output of the calibration generator to 60 MHz.
Termination
for P6021
calibration generator.
an amplitude of 5 V
in Table 8.
probe
Test oscilloscope
CH 1
pp
.
P6021
probe
Signal Out BNC connector on the rear of
Note: If using a Wavetek 9100, use the
Calibration
Coax cable
fixture
the instrument.
Service Information
Calibration
generator
Sinewave
Leveled
Output
17