P6021 Tektronix, P6021 Datasheet - Page 28

no-image

P6021

Manufacturer Part Number
P6021
Description
Test Probes 60MHZ, 250A 5 FT, w/term
Manufacturer
Tektronix
Datasheet

Specifications of P6021

Color
Black
Equipment Type
Test Probe Set
Accessory Type
AC Current Probe
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
For Use With
AC current measurements
Service Information
16
Table 7: Low frequency bandwidth measurements
Probe
Sensitivity
2 mA/mV
10 mA/mV
/
Leveled Sinewave
Output frequency
(Calibration
generator)
50 kHz
450 Hz
50 kHz
120 Hz
9. Measure and record the DMM output as M1 in Table 7.
10. Set the Leveled Sinewave Output of the calibration generator to 450 Hz.
11. Measure and record the DMM output as M2 in Table 7.
12. Calculate the low frequency bandwidth ratio:
13. Record the results in the test record on page 19 and compare the results
14. Set the P6021 termination sensitivity to 10 mA/mV.
15. Set the Leveled Sinewave Output of the calibration generator to 50 kHz and
16. Measure and record the DMM output as M3 in Table 7.
17. Set the Leveled Sinewave Output of the calibration generator to 120 Hz.
18. Measure and record the DMM output as M4 in Table 7.
19. Calculate the low frequency bandwidth ratio:
20. Record the results in the test record on page 19 and compare the results
NOTE. Dirty or worn mating surfaces between the transformer and the lid
degrade the low-frequency response. Clean them if necessary. See page 23 for
cleaning instructions.
M1
against the low frequency specification on page 19.
an amplitude of 5 V
against the low frequency specification on page 19.
M2
low frequency bandwidth ratio = M2
low frequency bandwidth ratio = M4
pp
.
M2
M1
M3
M1
M3
P6021 Instruction Manual
M4
M4
M3