DS3105DK Maxim Integrated Products, DS3105DK Datasheet - Page 18

no-image

DS3105DK

Manufacturer Part Number
DS3105DK
Description
Power Management Modules & Development Tools Demo Kit for DS3105 Demo Kit for DS3105
Manufacturer
Maxim Integrated Products
Datasheet

Specifications of DS3105DK

Lead Free Status / RoHS Status
Lead free / RoHS Compliant
_________________________________________________________________________________________DS3105DK
4.13 Configuration Scripts and Log File
4.13.1 Configuration Log File
Every write command issued by the software to the DS3105DK board is logged in file DS3105DKLog.mfg located
in the same directory as the software executable. This file can be viewed in Notepad by pressing the Log File
button in the upper-right corner of the main window. Command line option "-l <filepath>" can be used to cause the
software to write to a file other than DS3105DKLog.mfg.
4.13.2 Configuration Scripts
All or part of the text in the configuration log file can be copied to a text file with a .mfg file extension for use as a
configuration script. Configuration scripts are useful for quickly configuring the DS3105 without having to remember
all the required settings.
Two types of configuration scripts are possible: full and partial. A full configuration script can start with the DS3105
in its power-on default state and configure every aspect of the device to bring it to a desired state. To make a full
configuration script, run the software, uncheck the Demo Mode checkbox, initialize the device, configure the device
using the DK software fields, press the Log File button, and use File→Save As in Notepad to save a copy of the
entire log file to a different file name.
A partial configuration file only affects a subset of the DS3105 device settings. To make a partial configuration
script, press the Log File button to view the log file, press Ctrl-End to jump to the end of the file, and add to the
end of the file a carriage return or comment line (starting with a semicolon) to delimit the start of the desired
configuration. Then save and exit the log file. Next, configure the device using the DK software fields. Finally, view
the log file again, jump to the end, and copy everything from the delimiter to the end of the file into a new .mfg file.
To run a configuration script, press the Config Script button in the upper-right corner of the main window. In the
script window, type the path to the file or press the Browse button to navigate to the file.
Note that when the Demo Mode checkbox is unchecked, during the Initializing the DS3105 step, the software
runs configuration script startup.mfg located in the same directory as the software executable. The startup.mfg file
can be edited or replaced as needed to change the initial configuration of the device. Be aware, however, that the
section of the startup.mfg file labeled Required Initialization must be executed after device power-up or reset for
the DS3105 to operate correctly.