FLUKE-117 Fluke Electronics, FLUKE-117 Datasheet - Page 20

MULTMTR ELEC DMM W/NON CNT VOLT

FLUKE-117

Manufacturer Part Number
FLUKE-117
Description
MULTMTR ELEC DMM W/NON CNT VOLT
Manufacturer
Fluke Electronics
Series
VoltAlert™, 11xr
Type
Digital (DMM), Testerr
Datasheets

Specifications of FLUKE-117

Includes
Battery, Test Leads
Style
Handheld
Display Digits
3.75
Display Type
LCD, Bar Graph
Display Count
6000
Function
Voltage, Current, Resistance, Capacitance, Frequency
Functions, Extra
Continuity, Diode Test
Features
Backlight, Hold, Min/Max/Ave, Sleep
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Vendor undefined / RoHS compliant by exemption
Other names
2538815
614-1011
114, 115, and 117
Users Manual
To open the case for fuse replacement:
1. Remove the test leads from the Meter
2. Remove the Meter from its holster.
3. Remove two screws from the case bottom.
4. Separate the case bottom from the case top.
5. Remove the fuse from its holder and replace it with an
6. To re-assemble the Meter, first attach the case bottom
18
11 A, 1000 V, FAST fuse having a minimum interrupt
rating of 17,000 A. Use only Fluke PN 803293.
to the case top, then install the two screws. Finally,
insert the Meter into its holster.
Cleaning
Wipe the case with a damp cloth and mild detergent. Do
not use abrasives, isopropyl alcohol, or solvents to clean
the case or lens/window. Dirt or moisture in the terminals
can affect readings.