FLUKE-117 Fluke Electronics, FLUKE-117 Datasheet - Page 24

MULTMTR ELEC DMM W/NON CNT VOLT

FLUKE-117

Manufacturer Part Number
FLUKE-117
Description
MULTMTR ELEC DMM W/NON CNT VOLT
Manufacturer
Fluke Electronics
Series
VoltAlert™, 11xr
Type
Digital (DMM), Testerr
Datasheets

Specifications of FLUKE-117

Includes
Battery, Test Leads
Style
Handheld
Display Digits
3.75
Display Type
LCD, Bar Graph
Display Count
6000
Function
Voltage, Current, Resistance, Capacitance, Frequency
Functions, Extra
Continuity, Diode Test
Features
Backlight, Hold, Min/Max/Ave, Sleep
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Vendor undefined / RoHS compliant by exemption
Other names
2538815
614-1011
114, 115, and 117
Users Manual
22
AC Amps True-
rms
(45 Hz to 500 Hz)
DC Amps
Hz (V or A input)
Notes:
[1] All ac ranges except Auto-V LoZ are specified from 1 % to 100 % of range. Auto-V LoZ is specified from 0.0 V.
[2] AC Volts Hz is ac-coupled and specified from 5 Hz to 50 kHz. AC Amps Hz is dc-coupled and specified from 45 Hz
[3] >10 A unspecified.
Because inputs below 1 % of range are not specified, it is normal for this and other true-rms meters to display non-
zero readings when the test leads are disconnected from a circuit or are shorted together. For volts, crest factor of
V LoZ, AC mV, and AC amps are dc-coupled.
to 5 kHz.
[1]
Function
3 at 4000 counts, decreasing linearly to 1.5 at full scale. For amps, crest factor of 3. AC volts is ac-coupled. Auto-
[2]
6.000 A
10.00 A
20 A for 30 seconds max.,
10 minutes rest min.
6.000 A
10.00 A
20 A for 30 seconds max.,
10 minutes rest min.
99.99 Hz
999.9 Hz
9.999 kHz
50.00 kHz
[3]
[3]
Range
Table 1 Accuracy Specifications (cont.)
0.001 A
0.01 A
0.001 A
0.01 A
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
Resolution
([% of Reading] + [Counts])
Accuracy
1.5 % + 3
1.0 % + 3
0.1 % + 2
115, 117
115, 117
115, 117
Model