8846A/SU 120V Fluke Electronics, 8846A/SU 120V Datasheet - Page 65

PREC MLTMTR 6.5DIGI USB W/SFTWRE

8846A/SU 120V

Manufacturer Part Number
8846A/SU 120V
Description
PREC MLTMTR 6.5DIGI USB W/SFTWRE
Manufacturer
Fluke Electronics
Type
Digital (DMM)r
Datasheets

Specifications of 8846A/SU 120V

Includes
Test Leads
Style
Bench
Display Digits
6.5
Display Type
VFD, Dual
Display Count
Varies
Function
Voltage, Current, Resistance, Capacitance, Temperature, Frequency
Functions, Extra
Continuity, dB, Diode Test
Features
Memory, RS-232, USB Port
Ranging
Manual
Response
Average
Lead Free Status / RoHS Status
Contains lead / RoHS compliant by exemption
Other names
2675344
614-1003
Controlling System-Related Operations
Identifying Meter Errors
To copy the last or interrupted readings file from internal memory to an already-inserted
USB memory device:
1. Press M.
2. Press the INIT MEAS TO USB soft key.
The copy process will take several seconds.
To erase USB memory, refer to the “Media Memory” section in Chapter 1.
To clear internal memory contents:
3. Press M.
4. Press the MANAGE MEMORY soft key as shown below.
5. Press the ERASE MEMORY soft key.
6. If you are sure you want to clear all stored readings, all stored configurations, the
To check the amount of external memory available (8846A Only):
1. Press M.
2. Press the MANAGE MEMORY soft key.
3. Press the USB STATUS soft key.
When the Meter detects an error, the error annunciator (item 5 in Table 3-2) is
illuminated, and the beeper sounds. A list of possible Meter errors can be found in
Appendix B of this manual.
To read the error or errors:
1. Press I.
2. Press SYSTEM.
3. Press ERROR.
4. The first error, if there is more than one, is displayed. To read additional errors, press
If you want to erase all error messages without looking at each, press the CLR ALL soft
key.
user string and the hostname from internal memory, press the ERASE soft key. If
not, press the CANCEL soft key.
After a few seconds, the Meter displays the total external memory, the amount of
external memory used and free external memory.
NEXT.
F1
9.30011
F2
F3
F4
V
~
F5
Controlling System-Related Operations
Front-Panel Operation
caw062.eps
3-23